A liquid crystal lens can focus an endoscope without moving any lens element: it is a flat element whose optical power is changed by voltage, which makes it suitable for a distal lens that has no room for a motor and guides. The imaging lens of an endoscope sits at the very tip of the insertion tube, whose outer diameter is usually only a few millimeters. This space has to hold the observation window, the illumination fibers or LEDs, the objective, the image sensor and the wiring, and there is often no room left for a motor and guide rails. A conventional endoscope lens consists of fixed elements, so its depth of field is fixed and limited; when the focal plane needs to change, the traditional approach is to change the spacing between lenses in the lens group mechanically [1]. For flexible and capsule endoscopes, conventional focusing mechanisms are often impractical because of the limits on outer diameter and on lens displacement stroke [2].

The idea of a liquid crystal lens is to use a flat element whose liquid crystal layer is only a few to a few tens of micrometers thick and change its optical power by voltage, with no part of the lens group moving. This article examines whether this idea works in endoscopes and other miniature imaging devices. It first converts the focusing requirement into optical power, then looks at why a small aperture favors liquid crystal lenses, where in the lens the liquid crystal lens should go, and what limitations arise from drive voltage, polarization, body temperature and packaging; finally it compares the liquid crystal lens with miniature mechanical actuation and electrowetting liquid lenses.

The boundaries should be stated first. The endoscope and miniature imaging examples cited below are all laboratory research demonstrations; they assess optical feasibility and do not address clinical outcomes. Requirements specific to medical devices, such as electrical safety, biocompatibility, and cleaning, disinfection and sterilization, are discussed only qualitatively. Wherever there is no supporting literature, this article describes things as "application potential" or "example estimates".

Conventions in this article: optical power is P=1/fP=1/f, in diopters D (m−1\mathrm{m}^{-1}), positive for convergence; the object distance zz is the distance from the object to the lens, taken as positive; the liquid crystal layer thickness is dd, the birefringence Δn\Delta n and the clear aperture diameter DD, consistent with GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously. The origin of liquid crystal birefringence itself is explained in Liquid Crystal Birefringence: How Liquid Crystals Control the Phase and Polarization of Light.

Focusing needs in space-constrained systems

Endoscope tip with LC lens positions
Figure 1. Structure of an electronic endoscope tip and two integration positions for the liquid crystal lens (cross-section, schematic, not to scale; rays are not drawn with actual refraction). Illumination from the LEDs or fibers is directed onto the tissue, and the light reflected by the tissue or the fluorescence it emits is imaged through the observation window, the objective and the stop onto the image sensor. A is a liquid crystal lens at the stop, and B is a liquid crystal lens behind the observation window and in front of the objective; the two layers represent two liquid crystal layers with mutually perpendicular alignment. The drive wires leave through the cable together with the video and power wires; OD is the outer diameter of the insertion tube.

Depth of field of a fixed-focus lens

Endoscope objectives have short focal lengths and small pixels; for example, the endoscope used by Wang et al. had a lens focal length of 1.34 mm and a sensor pixel size of 1.4 μm [3]. As long as the aperture is not too large, such a lens stays sharp over a fairly wide range of object distances; that paper notes that the depth of field of endoscopes is typically around 30–40 mm [3].

On a millimeter scale the depth of field is not uniform, but this becomes clear in terms of vergence (the reciprocal of the object distance, 1/z1/z, also in D). Let the lens aperture be aa, the distance from the sensor to the lens s′s', and the permissible blur-spot diameter cc. When the vergence of the object deviates from the in-focus position by δV\delta V, the image-side convergence point moves accordingly, and the blur spot on the sensor has a diameter of a s′ ∣δV∣a\,s'\,|\delta V| (the derivation uses only the thin lens imaging formula [4]). Requiring this to be no larger than cc, the vergence range corresponding to the depth of field is

ΔV=2ca s′≈2caf=2cNf2\Delta V=\frac{2c}{a\,s'}\approx\frac{2c}{a f}=\frac{2cN}{f^2}

where N=f/aN=f/a is the f-number, and the approximation applies to relatively distant objects with s′≈fs'\approx f. This expression shows that, in vergence coordinates, the depth of field is roughly constant: the shorter the focal length, the larger the f-number and the larger the permissible blur spot, the wider the depth of field.

Example. Take a focal length f=1.34f=1.34 mm, f-number 4 (aperture 0.335 mm) and a permissible blur spot c=5.6 μmc=5.6\ \mu\mathrm{m} (4 pixels of 1.4 μm, comparable to the Airy disk diameter of about 5.4 μm at F/4 and 550 nm); this gives ΔV≈24.9\Delta V\approx24.9 D. Focused at 50 mm, the sharp range is about 30.8–132.9 mm; focused at 20 mm, about 16.0–26.6 mm; focused at 10 mm, about 8.9–11.4 mm; focused at 5 mm, only about 4.7–5.3 mm remains.

This estimate matches the order of magnitude of measured data. The system of Wang et al. had measured sharp ranges of 12–24 mm, 15–33 mm and 20–51 mm in three focus states [3]. In vergence, the widths of these three ranges are about 41.7 D, 36.4 D and 30.4 D, all of the order of 30–40 D. The same vergence width corresponds to a very narrow millimeter range at close distances; see Figure 3(b).

A "large depth of field" therefore holds only at medium and long distances. When the tip is brought close to the tissue to observe details at magnification, the depth of field shrinks to about a millimeter or less, and a fixed-focus lens must change its focus position. This is the main situation in which an endoscope needs focusing.

Requirements on a focusing device at the tip

Translated to the tip, these needs mean that a focusing device must meet at least the following requirements.

  • Size. The lateral size is limited by the outer diameter of the insertion tube, and the axial length directly adds to the length of the rigid tip section, affecting how well a flexible endoscope can bend through passages.
  • Power consumption and heat. The tip touches or is close to tissue, so its own heat generation must be minimal; capsule endoscopes are battery powered, so their power budget is even tighter.
  • Wire count. Each additional drive channel adds another wire running through the insertion tube.
  • Reliability. The tip has to withstand bending, impact and repeated cleaning and disinfection, and moving parts and seals are the weak points.

The potential advantages of a liquid crystal lens on these points are that the element is a flat plate with no moving parts, that electrically it is a small capacitor with very low power consumption, and that it may need only one or two drive lines. The following sections check quantitatively, one by one, the conditions under which these advantages hold.

Why a small aperture actually favors liquid crystal lenses

The optical path difference of a GRIN gradient-index liquid crystal lens comes from the radial variation of the refractive index within the liquid crystal layer, and the maximum optical path difference is no more than Δn d\Delta n\,d. A paraxial lens needs an optical path lead of PR2/2PR^2/2 at the edge of the aperture relative to the center (R=D/2R=D/2); equating the two gives the power limit of a single-layer lens (derivation in GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously):

Pmax⁡=8 Δn dD2P_{\max}=\frac{8\,\Delta n\,d}{D^2}

In large-aperture devices this relation is the biggest obstacle for liquid crystal lenses; Chen et al. call it the "power law" limitation of liquid crystal lenses, meaning that aperture can only be traded against optical power [5]. In an endoscope, however, the situation is reversed: the clear aperture is only one or two millimeters or even smaller, and with D2D^2 in the denominator, halving the aperture lets the same liquid crystal layer provide 4 times the optical power.

Example. For an optical path difference Δn d=4 μm\Delta n\,d=4\ \mu\mathrm{m} (for example Δn=0.2\Delta n=0.2 and d=20 μmd=20\ \mu\mathrm{m}), the limit is about 3.6 D for a 3 mm aperture, about 8 D for 2 mm, about 32 D for 1 mm and about 128 D for 0.5 mm. The same liquid crystal layer on a 10 mm aperture gives only about 0.32 D.

The response time changes in the same direction. After the voltage is removed, the relaxation time of the liquid crystal orientation is approximately τ≈γ1d2/(K11π2)\tau\approx\gamma_1 d^2/(K_{11}\pi^2), where γ1\gamma_1 is the rotational viscosity and K11K_{11} the splay elastic constant [6]. To reach optical power PP, the minimum cell gap needed is d=PD2/(8 δn)d=PD^2/(8\,\delta n), where δn≤Δn\delta n\le\Delta n is the refractive index difference actually used. Substituting gives

τ≈γ1K11π2⋅P2D464 δn2\tau\approx\frac{\gamma_1}{K_{11}\pi^2}\cdot\frac{P^2D^4}{64\,\delta n^2}

The response time is proportional to the fourth power of the aperture: halving the aperture makes the response 16 times faster at the same optical power.

Example. Take δn=0.2\delta n=0.2 and γ1/K11=4 ms/μm2\gamma_1/K_{11}=4\ \mathrm{ms}/\mu\mathrm{m}^2 (example values, the same as in GRIN Liquid Crystal Lenses). To obtain 20 D, a 1 mm aperture needs a cell gap of about 12.5 μm, with τ≈63\tau\approx63 ms; a 0.5 mm aperture needs about 3.1 μm, with τ≈4\tau\approx4 ms; a 2 mm aperture needs 50 μm, with τ\tau of about 1 s. To obtain 100 D, a 0.5 mm aperture needs a cell gap of about 15.6 μm, with τ≈99\tau\approx99 ms. These are all ideal lower bounds assuming δn=Δn\delta n=\Delta n.

Power limit and response time versus aperture
Figure 2. Power limit and response time at small apertures (calculated). (a) Power limit of a single-layer GRIN liquid crystal lens, Pmax⁡=8Δn d/D2P_{\max}=8\Delta n\,d/D^2, versus aperture DD, for optical path differences Δn d\Delta n\,d of 2, 4 and 21.5 μm; the open circle is the theoretical limit of about 43 D for the liquid crystal lens used by Wang et al. (Δn=0.43\Delta n=0.43, d=50 μmd=50\ \mu\mathrm{m}, 2 mm aperture), and the filled circle is the maximum positive power of 25.3 D reported in that paper [3]. (b) Relaxation time corresponding to the minimum cell gap required to reach target powers of 20, 50 and 100 D, with δn=0.2\delta n=0.2 and γ1/K11=4 ms/μm2\gamma_1/K_{11}=4\ \mathrm{ms}/\mu\mathrm{m}^2; the dashed line is 33 ms (the frame period at 30 frames per second), and the dots are the examples in the text.

Devices in the literature can be checked against this relation. The liquid crystal lens used by Wang et al. in an endoscope had a three-electrode structure in which one electrode had a 2 mm diameter hole, a 50 μm liquid crystal layer, and a liquid crystal with Δn=0.43\Delta n=0.43 at 546 nm [3]. By the expression above, its power limit is about 43 D. The paper reports that the positive power could be tuned continuously from 0 to 25.32 D and the negative power from 0 to −15.2 D [3]; the maximum positive power is about 59% of the theoretical limit, consistent with the point made in GRIN Liquid Crystal Lenses that the usable δn\delta n is clearly smaller than Δn\Delta n. This device also illustrates the cost: for a 50 μm liquid crystal layer, the example viscoelastic coefficient above gives a relaxation time on the order of seconds; the drive voltage for the positive power state in the paper was a combination of 90 Vrms and 30 Vrms [3].

Devices with smaller apertures have also been demonstrated. Hassanfiroozi et al. reported a multifunctional liquid crystal lens with dual-layer electrodes, only 1.42 mm in diameter, that can switch between 2D and 3D imaging modes and focus in both; to obtain an ideal gradient electric field, they coated the electrode with a high-resistance layer [7].

A small aperture also brings a structural convenience. GRIN Liquid Crystal Lenses points out that the field from a hole-patterned electrode penetrates from the hole edge toward the center over a distance of the same order as the electrode spacing, so a simple hole electrode is naturally suited to microlenses, and more complex structures such as high-resistance layers and multiple electrodes are needed only for large apertures. The apertures required in endoscopes fall right on the microlens side.

How the liquid crystal lens is integrated into a miniature lens

Converting optical power into focus distance

Start with the simplest case: a thin liquid crystal lens close to a thin objective, with a fixed sensor position. The powers of thin lenses in contact simply add, so the total power is P0+ΔPP_0+\Delta P, where P0P_0 is the power of the objective and ΔP\Delta P the power added by the liquid crystal lens. From the thin lens formula 1/z+1/s′=P0+ΔP1/z+1/s'=P_0+\Delta P [4], with s′s' unchanged,

1z=1z0+ΔP\frac{1}{z}=\frac{1}{z_0}+\Delta P

where z0z_0 is the focus distance when the liquid crystal lens adds no power. The focal length of the objective has dropped out of the expression: seen from object space, the power added by the liquid crystal lens adds directly to the vergence of the object, regardless of whether the objective focal length is 1 mm or 2 mm. Added positive power pulls the focus position closer, and negative power pushes it farther away; at ΔP=−1/z0\Delta P=-1/z_0 the lens focuses at infinity.

The data of Wang et al. can be used to check this relation. With no added power, their system focused at about 23 mm [3]. By the expression above, adding +11.4+11.4 D should focus at about 18.2 mm and adding −15.2-15.2 D at about 35.4 mm; the measured contrast peaks in the paper were at 18 mm and 38 mm [3]. The liquid crystal lens consisted of three glass substrates 0.7 mm thick [3]; the discrepancy at the far end may come from the difference between such a thick element and the thin lens model, and from the accuracy of determining the focus position from the contrast peak, but the order of magnitude and the trend agree. In that paper the liquid crystal lens was the frontmost element of the lens and the object distance was also measured from it; in this case the added power adds directly to the object vergence, independently of the spacing between the liquid crystal lens and the objective (for the reason, see "At the stop or in front of the lens" below).

Example. Pulling the focus position in from 50 mm to 10 mm requires an added 1/0.01−1/0.05=801/0.01-1/0.05=80 D; pulling it in to 5 mm requires 180 D; pulling it in from 30 mm to 15 mm requires about 33 D.

Focus conversion and depth of field
Figure 3. Relation between added power, focus distance and depth of field (calculated). (a) Focus distance calculated from 1/z=1/z0+ΔP1/z=1/z_0+\Delta P, for z0z_0 at infinity, 50 mm and 23 mm; the squares are the three states reported by Wang et al. (focus at about 38, 23 and 18 mm for ΔP=−15.2\Delta P=-15.2, 0 and +11.4+11.4 D) [3]. (b) Depth-of-field width versus focus distance: the solid line is the geometric model of the example in the text (ΔV=24.9\Delta V=24.9 D), the dots are the values listed in the text for 5, 10, 20 and 50 mm, and the squares are the widths of the three depth-of-field ranges measured by Wang et al. [3].

These numbers point to the most natural role for a liquid crystal lens in an endoscope. Switching between medium and long distances needs only a few tens of diopters, and a single depth of field is twenty to thirty diopters wide, so a few discrete focal states or a stretch of continuous tuning can stitch together the range that can be seen clearly. The endoscope system reported by Chen and Lin does exactly this: exploiting the ability of the liquid crystal lens to act as either a positive or a negative lens, it switches between three discrete optical powers and extends the clearly visible range to 12.4–76.4 mm, about twice that without the liquid crystal lens [1].

Close-up observation at magnification near the tissue is much harder. Pulling in from 50 mm to 5 mm requires 180 D, and by Figure 2(a) a liquid crystal layer with a 4 μm optical path difference reaches this level only for apertures below about 0.42 mm. This raises the question of integration position: where in the lens is the beam narrowest, while still letting the added power act fully on object space?

Comparison with mechanical focusing

If the same focusing were done by moving a lens element, how much stroke would it need? Using Newton's form of the imaging equation, xx′=f2xx'=f^2 (xx and x′x' measured from the object-side and image-side focal points, respectively) [4], changing the focus distance from z1z_1 to z2z_2 requires an image-side displacement of f2/(z2−f)−f2/(z1−f)f^2/(z_2-f)-f^2/(z_1-f).

Example. For a lens with a 1 mm focal length, pulling in from 50 mm to 10 mm requires a movement of about 0.09 mm, and to 5 mm about 0.23 mm; for a 2 mm focal length the figures are about 0.42 mm and 1.25 mm.

The required stroke is proportional to f2f^2 and becomes smaller as the focal length shortens; a stroke of a few hundred micrometers is not extreme in itself. The difficulty is that the tip has to accommodate an actuator, guides and end stops for it, keep the lens from decentering or tilting as it moves, and maintain this precision after repeated bending, impact and disinfection. A liquid crystal lens replaces "distance moved" with "optical power", and the parts that are hard to miniaturize in a mechanical solution, the actuator and the guide rails, simply do not exist in the liquid crystal solution. The two also scale differently: the mechanical stroke shrinks with f2f^2, whereas the added power required of the liquid crystal is determined only by the object distance and is independent of ff; whether it can be reached depends on the optical path difference of the liquid crystal layer and the width of the beam, as discussed in the next section.

At the stop or in front of the lens

The liquid crystal lens need not sit right against the objective. When it is placed inside the lens group, the effect of its added power on object space has to be converted according to the beam width. In the on-axis beam, the liquid crystal lens adds a defocus wavefront: the marginal ray has an extra optical path of ΔP y2/2\Delta P\,y^2/2 relative to the center, where yy is the height of the on-axis marginal ray at the liquid crystal lens. Under paraxial conditions this optical path difference is preserved as it propagates along the beam path; referred back to the entrance pupil, the corresponding change in object-space vergence ΔV\Delta V satisfies ΔV yEP2/2=ΔP y2/2\Delta V\,y_{\mathrm{EP}}^2/2=\Delta P\,y^2/2, i.e.

ΔV=ΔP(yyEP)2\Delta V=\Delta P\left(\frac{y}{y_{\mathrm{EP}}}\right)^2

where yEPy_{\mathrm{EP}} is the entrance pupil radius; the expression holds when the object distance is much larger than the length of the lens. Ray tracing leads to the same conclusion: after propagating a distance tt beyond a thin lens, the marginal ray height scales as 1−tV1-tV (VV being the vergence of the beam there), while a small change in vergence scales as (1−tV)−2(1-tV)^{-2} [4]; together these give exactly the square of the height ratio. When the liquid crystal lens is the first element, yy is simply the height of the object-space beam on it, ΔV=ΔP\Delta V=\Delta P, and the earlier conversion formula is recovered.

Combining this relation with the power limit Pmax⁡=8 δn d/D2P_{\max}=8\,\delta n\,d/D^2, and letting the aperture of the liquid crystal lens equal the on-axis beam diameter exactly (D=2yD=2y), the upper limit of the tunable object-space vergence is

ΔVmax⁡=8 δn dDEP2=8 δn d N2f2\Delta V_{\max}=\frac{8\,\delta n\,d}{D_{\mathrm{EP}}^2}=\frac{8\,\delta n\,d\,N^2}{f^2}

where DEP=f/ND_{\mathrm{EP}}=f/N is the entrance pupil diameter. Placing the liquid crystal lens where the beam is narrow raises its power limit, but its effect on object space shrinks by the square of the height ratio, and the two exactly cancel. What really matters is the optical path difference the liquid crystal layer can provide and the size of the entrance pupil, not the placement, provided the aperture of the liquid crystal lens is no larger than the on-axis beam.

Example. A lens with a 1.34 mm focal length at F/4 has an entrance pupil diameter of about 0.335 mm. With an optical path difference δn d=4 μm\delta n\,d=4\ \mu\mathrm{m}, ΔVmax⁡≈285\Delta V_{\max}\approx285 D, enough to cover the 180 D needed to go from 50 mm to 5 mm; with δn=0.2\delta n=0.2, 180 D requires a liquid crystal layer of about 12.6 μm, and the example viscoelastic coefficient gives τ≈65\tau\approx65 ms. The ratio of ΔVmax⁡\Delta V_{\max} to the depth-of-field width ΔV=2cN/f2\Delta V=2cN/f^2 above is 4 δn d N/c4\,\delta n\,d\,N/c, in which the focal length cancels; in this example it is about 11, meaning that the tuning range of this liquid crystal layer can stitch together roughly 11 depths of field.

The difference between positions lies in the off-axis field. Figure 1 marks two typical positions.

Position A: at the stop. Beams from all field points pass through the same area of the stop, so a liquid crystal lens aperture equal to the stop diameter suffices; the full optical path difference budget is available to every field point, and the added power is essentially uniform across the field. The costs are these: in a wide-angle endoscope, off-axis beams cross the stop at large angles, and since the effective refractive index of the liquid crystal layer depends on the angle of incidence, the added power and aberrations vary with field; the substrates of the liquid crystal lens take up axial space in a lens group that is already tight, and wires have to be routed inside the lens group.

Position B: behind the observation window, in front of the objective. This is the approach taken by Wang et al. and by Chen and Lin: the liquid crystal lens and a polarizer are attached in front of the endoscope lens [3][1]. It is the simplest retrofit and requires no changes to the existing lens group. Away from the stop, however, beams from different field points land on different areas of the liquid crystal lens, so its aperture DLCD_{\mathrm{LC}} has to cover the beams of the whole field, while each field beam occupies only a patch of diameter about DbeamD_{\mathrm{beam}}. The power limit is set by DLCD_{\mathrm{LC}}, but each beam actually receives only a fraction (Dbeam/DLC)2(D_{\mathrm{beam}}/D_{\mathrm{LC}})^2 of the defocus optical path, so most of the optical path difference budget is wasted, and the response time is that of a thicker liquid crystal layer. The lens of Wang et al. had a 2 mm aperture, while the endoscope lens they used had a focal length of only 1.34 mm [3]; for any f-number above about 0.7, the entrance pupil is smaller than this aperture. The tuning range of that lens, about 40 D, was limited by the power limit of the 2 mm aperture rather than by the entrance pupil.

Optically, then, the stop is the better position; for retrofitting and manufacturing, the front of the lens is more convenient. Wherever it is placed, the liquid crystal lens should be optimized together with the other elements as one surface of the lens group, rather than simply added on.

Microendoscopy: a different scale

Another class of miniature imaging devices works on a completely different scale. Bagramyan et al. built a microendoscopic probe from a fixed gradient-index (GRIN) rod lens and an electrically tunable liquid crystal lens for imaging biological tissue at different focal depths; by changing the drive frequency at constant voltage, the focus could be moved by about 74±3 μm74\pm3\ \mu\mathrm{m}, and imaging at different focal planes was tested in adult mouse brain slices and in the living mouse brain [8]. The task here is no longer to switch between tissue at 5 mm and at 50 mm, but to scan axially by a few tens of micrometers inside the tissue, the same kind of task as focal plane adjustment in a microscope; the relevant optical requirements are covered in Microscope Z-Scanning with Liquid Crystal Lenses: Fast Focusing Without Moving the Objective.

Drive voltage, electrical safety and heat

Electrically, a liquid crystal lens is approximately a capacitor, with the liquid crystal layer as the dielectric and transparent electrodes on either side. Liquid crystal devices are usually driven with AC voltages free of any DC component to avoid degradation from ion migration [6]. The current is mainly capacitive charging and discharging current, and the real power actually consumed is only a fraction of the apparent power VIVI.

Example. Assume an electrode overlap area of 2 mm×2 mm2\ \mathrm{mm}\times2\ \mathrm{mm}, a 20 μm liquid crystal layer and an average relative permittivity of 10, giving a capacitance of about 17.7 pF. Driven at 1 kHz and 5 Vrms, the current is about 0.56 μA and the apparent power about 2.8 μW; driving the same capacitance at 90 Vrms gives a current of about 10 μA and an apparent power of about 0.9 mW. The real power cannot exceed the apparent power.

In terms of power consumption and heat, the liquid crystal lens itself is therefore usually not the main heat source at the tip; what needs attention is the loss in resistive structures such as high-resistance layers, and the power consumption of the drive circuit itself.

The voltage level deserves more attention than the power. The 90 Vrms above is no exaggeration: the lens of Wang et al. needed such a voltage to obtain positive power [3]. For instruments inserted into the body, the higher the voltage at the tip, the heavier the burden on insulation, leakage current and isolation design; the relevant requirements are set by safety regulations and standards for medical electrical equipment and must be verified at the system level. There are two main ways to lower the voltage. One is to reduce the distance between the electrodes and the liquid crystal layer and to use high-resistance layers or materials with high dielectric anisotropy; GRIN Liquid Crystal Lenses discusses the relevant electrode structures. The other is to use frequency as the control variable. The liquid crystal autofocus lens demonstrated by Galstian et al. in a mobile phone camera tunes its optical power by changing the drive frequency at a fixed voltage, keeps spherical aberration below λ/10\lambda/10 over a range of up to 10 D, and demonstrated a relatively short autofocus convergence time [9]; the microendoscope of Bagramyan et al. also uses constant voltage with variable frequency, and the authors note that their design achieves lower-voltage, polarization-independent operation, improving the energy efficiency of the system [8]. Electrical control schemes of different liquid crystal devices are compared in Electrical Control of Liquid Crystal Devices: Response Time, Polarization Switches and Cascaded Switching.

The number of wires depends on the electrode structure. A single hole electrode plus a common electrode needs only two wires; a structure driven by two voltages needs three; orthogonal dual layers add no wires if they share the drive, but double the wire count if they need separate calibration.

Polarization: white light and fluorescence are both unpolarized

Endoscope illumination, tissue-scattered light and fluorescence are all essentially unpolarized. A single-layer GRIN liquid crystal lens acts on only one linear polarization component: the component parallel to the alignment direction sees the index profile and is focused, while the perpendicular component always experiences non_o and passes through unchanged (see GRIN Liquid Crystal Lenses). For unpolarized light, only half of the energy receives the added power, while the other half is still imaged at the original focal plane; the two images are superimposed on the sensor, one sharp and one blurred.

A polarizer can remove the unmodulated half, at the cost of losing at least 50% of the light. In the endoscope experiments of Wang et al., a polarizer was placed in front of the liquid crystal lens [3]. For white-light observation this means doubling the exposure time or the illumination power; for fluorescence imaging, where the signal is weak to begin with, losing half the light directly lowers the signal-to-noise ratio.

The standard way to avoid losing light is to stack two liquid crystal layers with mutually perpendicular alignment, so that each of the two orthogonal polarization components passes once through an identical lens. As early as 1985, Sato et al. stacked two liquid crystal Fresnel lenses with orthogonal optic axes to obtain a polarization-independent variable-focus lens [10]. Lin and Chen reported an electrically tunable-focusing liquid crystal lens that needs no polarizer and noted that the approach can also be used in imaging systems such as mobile phone cameras, pico projectors and endoscopes [11]; Chen et al. went on to realize a polarizer-free liquid crystal lens with an embedded multilayer structure and showed that increasing the number of liquid crystal layers can enlarge the aperture without reducing the tunable power [5].

Orthogonal dual layers raise a problem that needs particular attention in miniature lenses. The two liquid crystal layers are separated by a substrate, so the "lens planes" for the two polarization components are offset along the optical axis by a substrate thickness Δt\Delta t. By the height-ratio relation above, if the on-axis beam has different heights at the two layers, the same drive will have different effects on object space.

Example. Place the two layers behind a thin objective with a 1.34 mm focal length, in the beam converging to the image point of a distant object; the marginal ray height scales as 1−t/f1-t/f. With the layers 0.1 mm apart, the effect of the rear layer on object space is about (1−0.1/1.34)2≈0.856(1-0.1/1.34)^2\approx0.856 of that of the front layer, a difference of about 14%; with 0.3 mm spacing it is about 0.602, a difference of about 40%.

In systems with focal lengths of hundreds of millimeters this difference is negligible (GRIN Liquid Crystal Lenses estimates a focal difference of only about 0.1% of the focal length), but in an endoscope lens with a focal length of one or two millimeters it may exceed the depth-of-field tolerance. Mitigations include having the two layers share a very thin intermediate substrate; placing the dual layer in a space where the on-axis beam is nearly collimated and the beam height changes little with position; or driving and calibrating the two layers separately so that their effects on object space match.

A PB liquid crystal lens (see PB Lens Principles: How Geometric Phase Liquid Crystal Lenses Switch Focal Length Electrically) is at an even greater disadvantage here. It requires circularly polarized input of a single handedness, so unpolarized light again has to be polarized first; its optical power is proportional to wavelength, so chromatic aberration is pronounced in broadband white-light imaging; and its focal states are discrete. Its advantages, being thin, fast and available in large apertures, are not significant at the small apertures of endoscopes. For narrowband laser illumination or scanning imaging, PB lenses may still be worth considering, but this is an application potential that requires dedicated study.

Temperature: about 37 °C inside the body

The birefringence of a liquid crystal decreases as the temperature rises, and at the clearing point TcT_c the liquid crystal becomes an isotropic liquid and the birefringence vanishes. Based on Haller's approximation, Li et al. gave the temperature dependence of the birefringence as [12]:

Δn(T)=(Δn)0(1−TTc)β\Delta n(T)=(\Delta n)_0\left(1-\frac{T}{T_c}\right)^{\beta}

where TT and TcT_c are absolute temperatures in kelvin and (Δn)0(\Delta n)_0 and β\beta are material parameters. That paper gives fitted parameters for four liquid crystals at 589 nm [12], and Figure 4 plots the curves from 15 to 60 °C using these parameters.

Birefringence versus temperature
Figure 4. Birefringence of four liquid crystal materials versus temperature (calculated from the fitted parameters of Li et al. at 589 nm [12]). Vertical dashed lines mark 22 °C and 37 °C. The curves of 5CB and 5PCH drop to zero at their clearing points (about 33.5 °C and 52.8 °C); UCF-35 and MLC-6241-000 have clearing points of about 95 °C and 100 °C and decline only slowly over the range shown.

Example. From the expression above and the parameters in that paper, from 22 °C to 37 °C the birefringence of UCF-35 falls from about 0.369 to 0.347 (about −6.0%), that of MLC-6241-000 from about 0.086 to 0.082 (about −4.6%), and that of 5PCH from about 0.119 to 0.108 (about −9.6%); 5CB, with a clearing point of about 33.5 °C, is already an isotropic liquid at 37 °C and has lost its lens function completely.

These numbers yield two design requirements. First, the clearing point must be well above body temperature, with margin to cover the highest temperature the tip may experience; the closer the clearing point is to the operating temperature, the more steeply the birefringence changes with temperature. Second, even with a suitable material, the optical power drifts with temperature. The power of a GRIN lens is proportional to the usable refractive index difference, so at a fixed drive it changes roughly in proportion to the birefringence.

Example. If a drive state gives 40 D of added power at 22 °C, the −6.0% for UCF-35 implies about 2.4 D less at 37 °C, and the −4.6% for MLC-6241-000 about 1.8 D less.

A drift of a few diopters is smaller than the depth-of-field width estimated above (twenty to thirty diopters) and has little effect when several focal states are used to stitch together depth of field; but open-loop focusing based on lookup tables, or measurement applications that need to know the focus distance precisely, require temperature compensation. As temperature rises the rotational viscosity falls and the response speeds up, while the elastic constants and dielectric anisotropy change as well [6], so the whole relation between drive voltage and optical power shifts, and calibration should be done at the operating temperature. The tip temperature itself is not constant either: insertion from room temperature into the body, heat from the illumination source and irrigation fluid all change it, so closed-loop image-based focusing is more robust than open-loop lookup.

Comparison with miniature mechanical actuation and electrowetting liquid lenses

There are three main focusing options for space-constrained systems; Figure 5 sketches their structures.

Comparison of focusing mechanisms
Figure 5. Schematic structures of three focusing approaches. (a) Moving a lens: an actuator (such as a piezo or voice coil) moves the lens along the optical axis by Δ\Delta, which requires guides, stroke and mounting space. (b) Electrowetting liquid lens: a conductive liquid (aqueous phase) and an insulating liquid (oil phase) are sealed in a tube, the side-wall electrode is covered with an insulating layer, and the voltage changes the contact angle and curvature of the interface; the solid and dashed lines show the interface at two voltages V1V_1 and V2V_2. (c) GRIN liquid crystal lens: a flat liquid crystal cell in which a hole electrode and a high-resistance layer produce a radially varying electric field; the upper layer is aligned along xx and the lower along yy (the molecular projections indicate tilt increasing from center to edge), for use with unpolarized light; it is tuned by voltage VV and frequency ff.

Moving a lens. Optically this is the simplest: the elements themselves are glass or plastic lenses, so optical quality, chromatic aberration and polarization behavior are unaffected, and the focusing range can be large. The limitations are entirely mechanical: the actuator, guide rails and end stops must fit into the tip, the moving parts must be sealed, and everything must withstand bending and impact. Kuiper compared displacement-based and deformation-based focusing theoretically and concluded that in flexible and capsule endoscopes, conventional focusing mechanisms are unsuitable because of the limits on outer diameter and displacement stroke [2].

Electrowetting liquid lens. Berge and Peseux proposed controlling the focal length of a lens with an external voltage through the electrowetting effect [13]. Kuiper and Hendriks sealed two immiscible liquids in a tube, where electrowetting changes the curvature of the liquid–liquid interface and thus the focal length; the two liquids form a self-centered lens in the tube, and on this basis they built a miniature achromatic camera module [14]. Liquids are isotropic, so electrowetting lenses are inherently polarization independent; the optical power comes from the interface curvature and can vary over a large range. They have been demonstrated in endoscope settings: Seo et al. used MEMS processes to make a cylindrical electrowetting liquid lens whose lens module changed its focal length by about 166 mm over a voltage change of 11 Vrms, and after mounting it in a capsule endoscope they captured images of a pig colon at different focal lengths [15]; Kuiper also built a demonstrator of a liquid lens for endoscopes and verified its optical feasibility [2]. Their constraints come from the liquids themselves: some axial length is needed to hold the liquids and the curved interface, the interface shape depends on orientation, vibration and temperature, and the aperture is set by the tube diameter.

Liquid crystal lens. It is a flat plate whose thickness is set mainly by its substrates, with a liquid crystal layer only a few to a few tens of micrometers thick, and has no macroscopic liquid interface or moving parts. Its constraints have been calculated one by one above: the power limit is inversely proportional to the square of the aperture, the response time grows with the square of the cell gap, a single layer acts on only one linear polarization component, its parameters vary with temperature, and the drive voltage depends on the electrode structure.

Item Moving lens Electrowetting liquid lens GRIN liquid crystal lens
Moving parts Yes; guides and seals needed No macroscopic moving parts; the liquid interface deforms None; molecular orientation changes
Polarization Independent Independent Single layer is polarization dependent; orthogonal dual layers needed
Main limit on focusing range Stroke and space Interface curvature and liquid index difference Optical path difference and aperture squared
Main limit on speed Actuator and load mass Liquid motion Square of the liquid crystal layer thickness
Electrical characteristics Actuator power consumption Capacitive; voltage depends on dielectric layer and structure Capacitive; voltage depends on electrode structure
Space at the tip Large laterally and axially Axial length needed for the liquids Axial length mainly the substrates
Temperature sensitivity Relatively low Liquid properties and interface Birefringence, viscosity, clearing point

The comparison in the table is qualitative; the actual specifications of each approach depend on the device design and cannot be generalized by category. The potential advantages of liquid crystal lenses over liquid lenses are that they are thinner, more easily made as planar multilayer structures, and can reach large optical powers at apertures from micrometers to sub-millimeter; their disadvantages are polarization, temperature, and insufficient power beyond small apertures. Given that moving lenses and liquid lenses have already been studied extensively, the role of liquid crystal lenses in endoscopes is more likely to be a complementary option that is thinner and more power-efficient for specific apertures and focusing ranges than a universal replacement. The difference between focusing and zooming (changing magnification), and the general role of liquid crystal lenses in compact systems, are covered in What Is a Liquid Crystal Tunable Lens? Principles, Types and System Integration; autofocus in small camera modules is covered in Liquid Crystal Lens Autofocus: Electrical Focusing in Cameras and Machine Vision.

Where demonstrations end and application potential begins

The research cited above can be summarized by what it demonstrates.

  • Endoscope focusing and depth-of-field extension. Chen and Lin added a liquid crystal lens with both positive and negative power to an endoscope system and used three discrete optical powers to extend the clearly visible range to 12.4–76.4 mm [1]. Wang et al. mounted a 2 mm aperture liquid crystal lens and a polarizer in front of an endoscope lens; the focus position could be tuned continuously between 18 and 38 mm, and in synthetic aperture integral imaging the total depth of field was extended to 12–51 mm; the test objects were resolution charts and specimens, and the endoscope was mounted on a translation stage to capture multi-view images [3].
  • Small-aperture multifunctional liquid crystal lenses. The 1.42 mm liquid crystal lens of Hassanfiroozi et al. can switch between 2D and 3D modes and focus, and is aimed at 3D endoscopy [7].
  • Focal depth adjustment in microendoscopy. The GRIN rod lens plus liquid crystal lens probe of Bagramyan et al. achieved a focal shift of about 74 μm and imaged mouse brain slices and the living mouse brain [8].
  • Autofocus in miniature cameras. Galstian et al. demonstrated a liquid crystal autofocus lens in a mobile phone camera that kept spherical aberration low over a range of up to 10 D [9].
  • Variable focus in capsule endoscopes. Existing demonstrations use electrowetting liquid lenses, for example the images of a pig colon at different focal lengths by Seo et al. [15]. In the literature found for this article, there is no imaging demonstration of a liquid crystal lens in a capsule endoscope; this direction can currently be regarded only as application potential.

What these studies prove is that liquid crystal lenses can provide tens of diopters of tunable power at millimeter-scale apertures and can be integrated into the beam path of an endoscope or microendoscope to change the focus position. What they do not prove, and what this article does not judge, is stability after long-term use and repeated disinfection, image quality and handling in a real clinical environment, and whether there is any diagnostic benefit. The last two kinds of question need device-level design and clinical studies to answer.

Engineering challenges

Moving from laboratory demonstrations to usable instruments still requires solving the following kinds of problems.

Packaging and environmental tolerance. The liquid crystal layer, alignment layers, transparent electrodes and sealant are all enclosed between two substrates, and what contacts the tissue is the external observation window or lens barrel. The device has to withstand bending and impact at the tip, as well as the chemicals, moist heat or temperature cycling of cleaning, disinfection or sterilization. If a process exceeds the clearing point of the liquid crystal, the liquid crystal temporarily becomes isotropic; whether the alignment fully recovers after cooling and whether the package can withstand the process must be verified item by item. All external materials that contact the human body must meet biocompatibility requirements, and the liquid crystal itself must be guaranteed not to leak out over the whole service life.

System-level requirements for medical devices. Electrical safety, electromagnetic compatibility, risk management and software validation are all assessed at the system level, and the liquid crystal lens is only one component. A lower drive voltage and a simple wiring structure make the system design easier, but cannot replace these assessments.

Calibration and control. The relation between optical power and voltage or frequency is nonlinear, and it also varies with temperature and from device to device. A workable approach is factory calibration plus closed-loop image-based focusing: with image sharpness as feedback, the liquid crystal lens is treated as an actuator that responds relatively slowly but needs no mechanical positioning. During switching the phase profile is not an ideal parabola, so images captured during the transition need to be discarded or given a lower weight.

Optical quality. The aberrations of a liquid crystal lens change with optical power (see GRIN Liquid Crystal Lenses), and oblique incidence of off-axis beams in wide-angle lenses also causes field-dependent power and astigmatism. If the liquid crystal lens sits inside the lens group, it must be optimized together with the other elements, and the variation of its aberrations over the tuning range must be included in the tolerance analysis. Liquid crystal dispersion gives slightly different added powers at different wavelengths, so lenses used for both white light and near-infrared fluorescence need separate evaluation.

Manufacturing. Liquid crystal cells at the sub-millimeter scale require precise control of the cell gap, alignment and electrode pattern, and reliable electrode lead-out. If orthogonal dual layers, a thin intermediate substrate and mounting inside the lens group must all be accommodated at once, the device structure will be much more complex than a laboratory sample. Consistency in volume manufacturing directly determines how much calibration work each device needs.

Most of these challenges are engineering problems rather than obstacles in principle, but each needs an answer in a specific device and system design. The value of liquid crystal lenses in space-constrained imaging ultimately depends on whether they can provide a sufficient focusing range while satisfying the conditions of small aperture, low voltage, polarization independence and temperature stability at the same time.

References

[1] H.-S. Chen, Y.-H. Lin. An endoscopic system adopting a liquid crystal lens with an electrically tunable depth-of-field. Optics Express, 21(15): 18079–18088, 2013. DOI: 10.1364/OE.21.018079

[2] S. Kuiper. Electrowetting-based liquid lenses for endoscopy. Proceedings of SPIE, 7930 (MOEMS and Miniaturized Systems X): 793008, 2011. DOI: 10.1117/12.879133

[3] Y.-J. Wang, X. Shen, Y.-H. Lin, B. Javidi. Extended depth-of-field 3D endoscopy with synthetic aperture integral imaging using an electrically tunable focal-length liquid-crystal lens. Optics Letters, 40(15): 3564–3567, 2015. DOI: 10.1364/OL.40.003564

[4] E. Hecht. Optics, 5th ed. Pearson, 2017. ISBN 978-0-13-397722-6

[5] H.-S. Chen, Y.-J. Wang, C.-M. Chang, Y.-H. Lin. A polarizer-free liquid crystal lens exploiting an embedded-multilayered structure. IEEE Photonics Technology Letters, 27(8): 899–902, 2015. DOI: 10.1109/LPT.2015.2399932

[6] D.-K. Yang, S.-T. Wu. Fundamentals of Liquid Crystal Devices, 2nd ed. John Wiley & Sons, 2014. ISBN 978-1-118-75200-5. DOI: 10.1002/9781118751992

[7] A. Hassanfiroozi, Y.-P. Huang, B. Javidi, H.-P. D. Shieh. Dual layer electrode liquid crystal lens for 2D/3D tunable endoscopy imaging system. Optics Express, 24(8): 8527, 2016. DOI: 10.1364/OE.24.008527

[8] A. Bagramyan, T. Galstian, A. Saghatelyan. Motion-free endoscopic system for brain imaging at variable focal depth using liquid crystal lenses. Journal of Biophotonics, 10(6-7): 762–774, 2017. DOI: 10.1002/jbio.201500261

[9] T. Galstian, O. Sova, K. Asatryan, V. Presniakov, A. Zohrabyan, M. Evensen. Optical camera with liquid crystal autofocus lens. Optics Express, 25(24): 29945, 2017. DOI: 10.1364/OE.25.029945

[10] S. Sato, A. Sugiyama, R. Sato. Variable-focus liquid-crystal Fresnel lens. Japanese Journal of Applied Physics, 24(8A): L626, 1985. DOI: 10.1143/JJAP.24.L626

[11] Y.-H. Lin, H.-S. Chen. Electrically tunable-focusing and polarizer-free liquid crystal lenses for ophthalmic applications. Optics Express, 21(8): 9428–9436, 2013. DOI: 10.1364/OE.21.009428

[12] J. Li, S. Gauza, S.-T. Wu. Temperature effect on liquid crystal refractive indices. Journal of Applied Physics, 96(1): 19–24, 2004. DOI: 10.1063/1.1757034

[13] B. Berge, J. Peseux. Variable focal lens controlled by an external voltage: An application of electrowetting. The European Physical Journal E, 3(2): 159–163, 2000. DOI: 10.1007/s101890070029

[14] S. Kuiper, B. H. W. Hendriks. Variable-focus liquid lens for miniature cameras. Applied Physics Letters, 85(7): 1128–1130, 2004. DOI: 10.1063/1.1779954

[15] S. W. Seo, S. Han, J. H. Seo, Y. M. Kim, M. S. Kang, N. K. Min, W. B. Choi, M. Y. Sung. Microelectromechanical-system-based variable-focus liquid lens for capsule endoscopes. Japanese Journal of Applied Physics, 48(5R): 052404, 2009. DOI: 10.1143/JJAP.48.052404