A microscope can image only a very thin layer of the sample sharply at a time. To obtain the three-dimensional structure of cells, tissue sections or brain regions in vivo, the focal plane has to be moved layer by layer along the optical axis, an image recorded at each layer, and the resulting stack of images (z-stack) assembled into volume data; this is microscope Z-scanning. A liquid crystal lens can change the layer while both the objective and the sample stay still: placed in the beam path behind the objective, it uses a voltage to change the vergence of the beam, and the focal plane moves along the optical axis accordingly. For fixed samples, slower layer switching only costs more time; for living samples, the slower the switching, the more the images of different layers differ in acquisition time, and fast physiological events are split across different moments in the volume data.
The conventional approach is to move the objective or the sample: a piezo positioner drives the objective up and down, or the Z axis of the stage moves the sample. These methods are accurate and mature, but in every case the part that moves is close to the sample and has appreciable mass. Grewe et al. pointed out that in in vivo two-photon calcium imaging, three-dimensional scanning with galvanometer lateral scanning plus a piezo-driven objective reaches sampling rates of at most about 10 Hz, limited mainly by the inertia of the moving objective; the reciprocating vibration of the objective also makes simultaneous electrophysiological recordings hard to keep stable [1]. Botcherby et al. likewise noted that the speed of the piezo approach is limited by the mass of the objective or sample, and that it makes imaging of delicate samples such as live cell cultures difficult [2]. In light-sheet microscopy, Fahrbach et al. stressed that moving the sample limits speed and precision and may also damage the sample [3].
Another approach keeps both the objective and the sample still and instead changes the convergence or divergence of the beam in a part of the beam path farther from the sample, so that the objective itself places the focus at different depths. Electrically tunable lenses (elastic-membrane or electrowetting liquid lenses, liquid crystal lenses and so on) are the components that do this. This article discusses how liquid crystal tunable lenses are used in such systems: the quantitative relation between added power and focal shift, where the device should be placed, how confocal, two-photon and widefield systems each integrate it, what optical performance the liquid crystal lens must meet, and where it stands relative to other fast axial scanning approaches. The focusing principles of liquid crystal lenses themselves are covered in GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously and PB Lens Principles: How Geometric Phase Liquid Crystal Lenses Switch Focal Length Electrically. The subject here is focusing, with the magnification kept constant; the difference between focusing and optical zoom is explained in What Is a Liquid Crystal Tunable Lens? Principles, Types and System Integration.
Conventions in this article: optical power is , in diopters D (i.e. ), positive for convergence; the power added by the liquid crystal lens is written . The axis runs along the optical axis, positive from the objective into the depth of the sample, so means the focus moves deeper, away from the objective. The effective focal length of the objective is (the tube lens focal length divided by the nominal magnification), its numerical aperture is , and the refractive index of the immersion medium is . All examples assume a 200 mm tube lens, and the refractive indices of water and oil are taken as 1.333 and 1.518; these are example values.
What axial scanning has to deliver
Different microscopy modalities place different demands on axial scanning, so the requirements are worth stating first.
A widefield fluorescence microscope captures the whole field of view at once, but fluorescence from out-of-focus layers also reaches the detector and forms a blurred background. A confocal microscope places a pinhole conjugate to the focus in the detection path, which passes only the light emitted near the focus and blocks most out-of-focus light, giving it optical sectioning capability [4]. This also means that each time a confocal microscope changes layers, the excitation focus and the conjugate position of the pinhole must move together.
A two-photon microscope excites with femtosecond near-infrared pulses, and fluorescence is generated only at the laser focus; scanning the focus laterally produces a two-dimensional image, and moving it along the optical axis produces three-dimensional data [1]. Because the origin of the fluorescence is already confined by the excitation focus, no pinhole is needed on the detection side; this is key to choosing where to place the liquid crystal lens, as discussed later.
For in vivo functional imaging, what matters is often not a continuous dense z-stack but rapid back-and-forth switching between a few layers. In mouse barrel cortex, Grewe et al. alternated between two layers 40 μm apart to record the activity of about 40 neurons, at a total sampling rate of 20–30 Hz [1]. Fahrbach et al. recorded 17 planes of a beating zebrafish heart at 510 frames per second, equivalent to 30 volumes per second [3].
In summary, an axial scanning device has to meet several requirements at the same time:
- Range. From tens of micrometers to several hundred micrometers, depending on sample thickness and objective.
- Speed. The layer switching time should be comparable to or shorter than the acquisition time of one layer; otherwise most of the volume rate is consumed by switching.
- Repeatability. After many round trips, the error in returning to the same layer should be smaller than the axial resolution. Grewe et al. reported an axial positioning error below 1 μm for a tunable lens over 2000 back-and-forth switches [1].
- Image quality. Switching layers should not noticeably degrade resolution, and ideally should not change magnification either, so that images of different layers can be overlaid and compared directly.
- Transmission. Fluorescence signals are usually weak, and every bit of light lost in the detection path directly lowers the signal-to-noise ratio.
Focal shift and added power
At the back focal plane of the objective
Modern microscope objectives are designed for infinity conjugates: with the sample at the front focal plane, the objective emits a collimated beam, which the tube lens then images. The aperture stop of the objective is at its back focal plane (BFP), so the object-space chief rays are parallel to the optical axis and the objective is telecentric in object space [5].
Suppose a thin lens of power is placed at the back focal plane and a collimated beam arrives from the tube lens side. After this lens, the beam acquires a vergence , as if heading for a point beyond the back focal plane. For the objective this point is an "object", at a distance from the objective's back focal point. The focal length of the objective is on the air side and on the immersion side. By Newton's imaging equation, the distance from the object to one focal point and the distance from the image to the other focal point satisfy , so the image point lies at a distance from the front focal point of
Now consider the direction. A converging beam () aims at a point beyond the back focal point, on the objective side, which is equivalent to an object lying "inside" the focal point; the corresponding image point then lies inside the front focal point, i.e. closer to the objective. With the axis convention of this article this is the negative direction, so
A converging added power pulls the focus toward the objective, and a diverging added power pushes it deeper into the sample. Martínez-Corral et al. used the same geometric relation to show that placing a lens at the back focal plane of the objective leaves the focal length and the back focal plane position of the combined system unchanged; only the front focal plane moves, by an amount proportional to the power of this lens, and a converging lens moves it toward the objective [5].
Three points about this result are worth noting. First, the focal shift is proportional to the square of the objective focal length, so low-magnification objectives respond much more strongly to the same . Second, the immersion medium magnifies the focal shift by a factor of . Third, in the paraxial approximation the relation is strictly linear, with no term, because the added lens sits exactly at the focal plane; as shown later, the linearity breaks down once the added lens leaves that plane.
Relaying to an accessible position
The back focal plane usually lies inside the objective barrel, where a device cannot necessarily be inserted directly. In the objective used by Grewe et al., the stop could not be accessed directly [1]; to align precisely with the optical axis without reducing the effective aperture, Martínez-Corral et al. added an afocal relay between the objective and the tube lens and placed the tunable lens at a plane conjugate to the aperture stop [5].
The usual relay is a 4f system: the front focal plane of the first lens (focal length ) coincides with the back focal plane of the objective, and the back focal plane of the second lens (focal length ) is the conjugate of the back focal plane; the liquid crystal lens is placed there, as in Figure 1. The pupil magnification is written
where and are the beam diameters at the objective's back focal plane and at the liquid crystal lens, respectively. means that the beam at the liquid crystal lens is smaller than the objective pupil.
The relay magnifies heights by and reduces angles to . The vergence of a beam can be thought of as the ratio of the marginal ray angle to its height, so on its way from the liquid crystal lens to the back focal plane it is reduced to of its value, and the of the liquid crystal lens is equivalent to at the back focal plane [5]. Substituting into the expression above:
The larger , the smaller the focal shift per diopter from the same liquid crystal lens, but also the smaller the aperture it needs.
By the sine condition, the pupil radius of the objective is proportional to the sine of the object-space aperture angle [2], and the pupil diameter is . The table below lists several common objectives for a 200 mm tube lens.
Example. Focal shift of each objective for D of added power (calculated, paraxial formula):
| Objective (example) | (mm) | (mm) | Focal shift at | Focal shift at |
|---|---|---|---|---|
| 10×/0.30 air | 20 | 12.0 | ∓400 μm | ∓100 μm |
| 20×/0.75 air | 10 | 15.0 | ∓100 μm | ∓25 μm |
| 25×/1.05 water immersion | 8 | 16.8 | ∓85.3 μm | ∓21.3 μm |
| 40×/0.80 water immersion | 5 | 8.0 | ∓33.3 μm | ∓8.3 μm |
| 60×/1.40 oil immersion | 3.33 | 9.33 | ∓16.9 μm | ∓4.2 μm |
| 100×/1.40 oil immersion | 2 | 5.6 | ∓6.1 μm | ∓1.5 μm |
In the table, "∓" means that D gives a negative focal shift (focus closer to the objective) and D a positive one. For the same 1 D, the focal shift of the 10× objective is about 66 times that of the 100× oil objective. Figure 2(a) plots the – lines of several objectives at .
Expressing focal shift as optical path difference
The formula above involves three quantities, , and ; for a liquid crystal lens there is a more direct way to write it. The beam radius at the liquid crystal lens is , and a thin lens of power provides an optical path difference between the beam edge and the center of . Substituting into the focal shift formula, and cancel together:
This expression says that moving the focus by in a medium of index requires the pupil edge to have an extra optical path of relative to the center, which is exactly the size of the paraxial defocus wavefront. The relay magnification and the aperture of the liquid crystal lens only decide whether this optical path difference is provided as "large aperture, small power" or as "small aperture, large power"; they do not change the optical path difference itself.
For liquid crystal lenses this point is crucial. The optical path difference of a GRIN liquid crystal lens comes from a refractive index difference, and a single layer can provide at most (GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously); so the focal shift range it can provide is set first of all by the optical path difference and has little to do with the choice of aperture or relay magnification.
Example. For a 40×/0.80 water immersion objective with , 1 D corresponds to an optical path difference at the beam edge of ; substituting into the expression above gives , in agreement with the table. If a scan of only is needed, the required optical path difference is , about 4.4 times the 550 nm wavelength.
The in the formula is the numerical aperture actually used by the beam. When the beam does not fill the objective pupil, the effective is smaller and the required optical path difference falls with . In the two-photon system of Grewe et al., limited by the scanner angle, the pupil of the 0.8 NA objective was not filled, and simulations gave an effective excitation NA of about 0.5 at the zero position [1].
Why a conjugate plane preserves magnification
Object-space telecentricity means that the chief ray from every object point is parallel to the optical axis in object space; after entering the objective it crosses the optical axis at the back focal plane, and after the relay it crosses the axis again at the conjugate of the back focal plane, as shown by the red line in Figure 1. The liquid crystal lens sits exactly there, so the chief ray passes through the lens center () and is not deflected at all by the added power. Since the chief ray path is unchanged, the image position of each object point on the detector is unchanged; whatever the value of , the magnification stays the same, and the new focal plane is still telecentric in object space. On the other hand, the beam diameter at the back focal plane is unchanged, so the effective aperture of the objective is unchanged too; as long as the liquid crystal lens does not clip the beam, the resolution does not change with the layer [5]. Martínez-Corral et al. confirmed this with an electrowetting liquid lens placed at a plane conjugate to the aperture stop: over the whole scan range, the full width at half maximum of the lateral point spread function varied by less than 3% [5].
If the liquid crystal lens is displaced along the optical axis by a distance from the conjugate plane, the chief ray crosses the lens at a height equal to times the chief ray angle and is deflected. A paraxial ray transfer matrix calculation shows that the relative change in magnification after this displacement is about ( means the liquid crystal lens is farther from the objective than the conjugate plane), and the focal shift becomes
so the linearity is lost. Here both and are measured on the liquid crystal lens side; when converted to the back focal plane, is multiplied by and divided by , so their product is unchanged.
Example. With mm and D, the magnification changes by about ; for a field of view 500 μm in diameter, the image points at the edge of the field move by about 2.5 μm (referred to object space).
Grewe et al. mounted the tunable lens and a negative offset lens directly in front of the objective, not at a plane conjugate to the stop, so the system was no longer telecentric: as the focus moved toward the objective, the field of view size changed with a slope of about 13% increase per 100 μm; simulations also showed that, with the pupil filled, the effective NA fell from 0.8 to about 0.5 with focal shift [1]. In their discussion they suggested re-imaging the pupil with a 4f system made of two achromatic lenses so that the system would be telecentric at the zero position [1]. Fahrbach et al. used a telecentric arrangement in light-sheet microscopy, which solved the problem of magnification changing from layer to layer [3][5].
Spherical aberration at high numerical aperture
The paraxial formula tells where the focus moves, but not what the image quality is once it gets there. This question is acute for high-NA objectives.
Microscope objectives are designed to satisfy the sine condition, which guarantees aberration-free imaging of points in a plane perpendicular to the optical axis; aberration-free imaging of points distributed along the optical axis requires the Herschel condition instead, and except at special magnifications the two conditions cannot be satisfied simultaneously [2]. An objective is therefore ideal only for its design focal plane; when layers are changed by altering the vergence of the incident beam, the farther the focus is from the design focal plane, the larger the spherical aberration [2].
This can be written quantitatively. Let an object point lie a distance from the focal plane along the optical axis. By the sine condition, the ray direction corresponding to normalized radius in the objective pupil satisfies , where . The optical path difference this object point produces in the pupil (relative to the center) is [2]
To move the focus to this point, the pupil must supply the opposite wavefront. A thin lens at a pupil conjugate plane can supply only the term, i.e. the first term, which is the origin of the paraxial formula above; it cannot supply the and higher terms, and these are the spherical aberration introduced by refocusing. The larger the numerical aperture, the closer is to 1 and the larger the higher-order terms.
In actual imaging, the focus automatically settles where the residual wavefront error is smallest, i.e. where the term and a constant term best fit ; only what remains is the true residual aberration. We performed this least-squares fit for a uniformly illuminated circular pupil, and the results are plotted in Figure 2(b). The rms residual error is proportional to . By the Maréchal criterion, an rms wavefront error no larger than gives a Strehl ratio of about 0.8, and imaging is close to diffraction-limited [6].
Example. Take ( nm). The maximum focal shift that can be achieved with parabolic phase focusing alone while remaining close to diffraction-limited is about for an NA 0.5 air objective, for NA 0.8 water immersion, for NA 1.0 water immersion and only for NA 1.4 oil immersion. The limit is proportional to wavelength; at 0.92 μm, commonly used for two-photon excitation, the limit for NA 0.8 water immersion is about .
Keeping only the term, the rms residual can be approximated as , which is accurate enough for small NA; for NA 1.4 oil immersion the higher-order terms contribute heavily, and the exact value is about 2.8 times this approximation.
The same fit also yields a result that is easy to overlook: at high NA, the actual shift of the best focus is smaller than the paraxial formula predicts. The reason is that the fitted coefficient is larger than the paraxial coefficient , so the same added power "buys" a shorter focal shift. By calculation, the ratio of the actual focal shift to the paraxial prediction is about 0.98 for NA 0.3 air, 0.90 for NA 0.8 water immersion, 0.84 for NA 1.0 water immersion and 0.72 for NA 1.4 oil immersion. In high-NA systems, therefore, the relation between voltage, optical power and focal shift must be calibrated by measurement rather than taken directly from the paraxial formula.
These are estimates for an ideal objective. The internal aberrations of a real objective are finely balanced for collimated input, and this balance is broken when a converging or diverging beam enters [1]. Simulations of a 40×/0.8 objective by Grewe et al. showed that when the focus moves toward the objective, the loss of resolution from the reduced effective NA masks the added aberrations, so relatively large shifts are possible; when the focus moves deeper, the effective NA increases, aberrations become the limit, and the usable focal shift range is much smaller [1]. Koukourakis et al. used a tunable lens for axial scanning in a confocal microscope and found that the defocus and spherical aberration introduced by focusing moved the observation plane away from the best focus, degrading the axial resolution by a factor of about 2.4; they added a second tunable lens to the detection path to pull the observation plane back to the best focus, keeping the axial resolution uniform during the scan [7].
The fundamental way around this limit is remote focusing: a second objective replicates the sample space at a remote location with a magnification of ; such a system can satisfy both conditions at once, and layers are changed by moving a mirror at the remote location [2]. Botcherby et al. used this method to achieve aberration-free refocusing over a 70 μm range with a 1.4 NA objective [8].
For liquid crystal lens applications, these results imply two things. With low or moderate NA, or when the beam does not fill the pupil, a liquid crystal lens at a pupil conjugate plane can maintain image quality over a range of tens to hundreds of micrometers; with high-NA objectives, a device that provides only a parabolic phase can do no more than fine adjustments on the micrometer scale. The phase profile of a liquid crystal phase device can be shaped with multiple electrodes (GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously), so in principle a term could be added together with the defocus to precompensate spherical aberration, with one set of voltages for each target focal shift. This is an application potential; in the literature searched for this article, no experimental report of doing this with a liquid crystal lens in a microscope was found.
Example. For NA 1.4 oil immersion and , the term of at the pupil edge is , about 5 times the 550 nm wavelength.
Integrating into confocal, two-photon and widefield systems
The part of the beam path in which the liquid crystal lens is placed determines which light passes through it, and therefore the polarization and spectral conditions it must meet. Figure 3 compares several integration schemes.
Confocal: excitation and detection must move together
The confocal pinhole is conjugate to the excitation focus [4]. If only the excitation focus moves while the conjugate point of the detection does not, the pinhole blocks precisely the light emitted from the new focus. The liquid crystal lens must therefore be placed in the path shared by excitation and detection, usually at a pupil conjugate plane between the scanner and the objective, as in Figure 3(b). The excitation light passes through it once on the way in and the fluorescence passes through it again on the way back, so their foci move together. The scanner itself sits at a pupil conjugate plane, and the relay between the scanner and the objective creates further conjugate planes; the liquid crystal lens can use these positions, with an additional relay if necessary.
This arrangement raises two issues related to the liquid crystal.
The first is polarization. Even with linearly polarized laser excitation, the fluorescence emitted by randomly oriented fluorophores is only partially polarized, and molecular rotation depolarizes it further [9]. A single-layer GRIN liquid crystal lens acts on only one linear polarization component, while the other component sees only a uniform flat plate (GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously). Placed in the shared confocal path, this has two consequences. Of the fluorescence emitted from the new focal plane, the half not acted on by the lens does not return to the pinhole, so the signal is halved. Of the fluorescence emitted from the original focal plane, that same polarization component is still focused exactly on the pinhole. With single-photon excitation the original focal plane lies within the excitation cone and also fluoresces, so signal from the original focal plane mixes into the image of the new focal plane and the optical sectioning capability deteriorates. In the shared path of a fluorescence confocal microscope, the liquid crystal lens therefore needs a polarization-independent structure, such as two liquid crystal layers with orthogonal alignment. Lin et al. listed polarization dependence as one of the three main problems hindering liquid crystal lens applications and discussed design principles for polarization-independent liquid crystal lenses [10].
The second is dispersion. The excitation and fluorescence wavelengths differ, and the optical power of the liquid crystal lens is not exactly the same at the two wavelengths, so the excitation focus and the detection conjugate point separate. For a PB liquid crystal lens, the optical power is proportional to wavelength (PB Lens Principles: How Geometric Phase Liquid Crystal Lenses Switch Focal Length Electrically), so the dispersion is deterministic; for the GRIN type, the dispersion comes from the dispersion of the liquid crystal refractive indices, is milder and has to be calculated for the material used.
Example. For 488 nm excitation and 520 nm detection, the ratio of the PB lens powers at the two wavelengths is . If a 20 μm focal shift is designed at 488 nm, the conjugate point on the fluorescence side moves about 1.3 μm farther, already of the same order as the axial resolution of high-NA confocal microscopy. The fluorescence itself also has a bandwidth: if it covers 500–540 nm, the relative change in power across the band is about 7.7%.
Reflection confocal microscopy has neither problem: the returning light has the same wavelength as the illumination and largely preserves its polarization. Khan and Riza reported a no-moving-parts axial scanning confocal microscope that used a liquid crystal lens 5 mm in diameter with a 0.65 NA objective to move a 633 nm focal spot electrically and continuously over a 48 μm range; they measured a 3 dB axial resolution of 3.1 μm and imaged an indium phosphide dual square optical waveguide sample 2.3 μm wide and 2.3 μm high; using the fine analog electrical control of the liquid crystal lens, they also demonstrated an axial resolving capability of 270 nm [11]. Estimated with the optical path difference relation above, if the beam filled 0.65 NA, the 48 μm span would correspond to an optical path difference of about 10.1 μm, about 16 times the 633 nm wavelength; this is only an estimate from the formulas in this article, not a device parameter given in the paper.
As for speed, point-scanning confocal microscopy must complete a full lateral scan for each layer, so layer changes are relatively infrequent. On this basis, Martínez-Corral et al. argued that axial scanning speed is not the main issue for confocal microscopy [5]. This judgment applies to galvanometer scanning with long frame times; when a resonant scanner images at tens of frames per second, the switching time becomes a bottleneck again, as shown in the timing analysis below.
Two-photon: only the excitation light passes through the liquid crystal lens
Two-photon excitation generates fluorescence only at the focus [1], and detection is usually non-descanned: behind the objective, a dichroic mirror reflects the fluorescence directly to the detector, and it does not return to the scanner. Only the excitation focus has to move, so the liquid crystal lens can be placed at a pupil conjugate plane between the laser and the scanner, as in Figure 3(c), and the fluorescence never passes through it. Discussing improvements, Grewe et al. also noted that once the pupil is re-imaged with a 4f system, the fluorescence is reflected to the detector by a dichroic mirror near the objective and no longer passes through the tunable lens, which improves detection efficiency [1].
This arrangement neatly avoids the main weaknesses of liquid crystal lenses in a fluorescence path:
- Polarization. Femtosecond lasers are linearly polarized; aligning the polarization direction with the liquid crystal alignment direction makes a single-layer liquid crystal lens sufficient, with no need for orthogonal dual layers. If a PB liquid crystal lens is used, a quarter-wave plate converts the light to circular polarization.
- Dispersion. The excitation spectrum is relatively narrow, so dispersion affects the excitation focus much less than in a fluorescence path; PB lenses or phase-wrapped diffractive liquid crystal lenses may also be usable.
- Optical path difference. Two-photon imaging often does not fill the objective pupil, so the effective NA is smaller; the optical path difference required for a given focal shift falls with , and so does the spherical aberration.
What needs to be checked is the phase retardation and transmission of the liquid crystal lens at the near-infrared operating wavelength and its tolerance of high-peak-power pulses, all of which have to be evaluated for the specific material and structure. As a reference, the elastic-membrane tunable lens plus offset lens combination used by Grewe et al. had a total transmission of about 81% at the 850 nm excitation wavelength and about 87% in the visible range [1].
Example. For a 40×/0.80 water immersion objective with and a D PB liquid crystal lens in the excitation path, the paraxial formula gives switching between three focal planes spaced about 16.7 μm apart ( D, 0 and D; this requires a three-state structure made of a polarization switch and an active PB lens; see PB Lens Principles: How Geometric Phase Liquid Crystal Lenses Switch Focal Length Electrically). This combination is currently an application potential; in the literature searched for this article, no report of actually using a PB liquid crystal lens to change layers in a two-photon microscope was found.
The experiments of Grewe et al. provide a performance reference for this scenario: with an elastic-membrane tunable lens in the excitation path of a 40×/0.8 water immersion objective and an optimized drive signal, layer switching took about 15 ms, the maximum axial range was 0.7 mm, and in vivo calcium imaging of two layers at 20–30 Hz was achieved [1].
Widefield and light-sheet: the liquid crystal lens in the detection path
In a widefield microscope the "focal plane" is defined by the detection side, so the tunable lens can only be placed at a pupil conjugate plane between the objective and the tube lens [5], and all the fluorescence from the sample has to pass through it. This is the most demanding scenario for a liquid crystal lens: it must be polarization independent, have the same optical power across the whole fluorescence band, and offer very high transmission.
Martínez-Corral et al. compared two tunable lenses in the same telecentric widefield microscope. The electrowetting liquid lens gave axial scanning with constant magnification and unchanged resolution, with a response time of about 9 ms [5]. Their home-made liquid crystal lens used E7 liquid crystal in a structure of two layers, each with a 60 μm cell gap, to obtain a larger optical power; it did move the object plane axially, but its point spread function was markedly worse than that of the original microscope and the liquid lens, and the image quality and contrast were unacceptable; the response time ranged from 200 to 500 ms depending on the liquid crystal layer thickness [5]. The authors considered liquid crystal lens technology to be still at an early stage, with thinness and light weight as its main advantages [5].
In a light-sheet microscope the detection focal plane must follow the light sheet. Fahrbach et al. used an electrically tunable lens for remote focusing, without moving the sample or the objective, and imaged the zebrafish heart at 30 volumes per second [3]. At such frame rates the switching time has to be on the millisecond scale, and together with the requirements of polarization independence and broadband operation, this is hard for current liquid crystal lenses to meet.
Optical performance the liquid crystal lens must meet
Optical path difference sets the range, layer thickness sets the speed
Chaining the relations above together reveals the core trade-off of a GRIN gradient-index liquid crystal lens in a microscope. For a required axial scan span , the liquid crystal lens must provide an optical path difference span of . With a continuous phase profile and a usable refractive index difference (no more than ), the liquid crystal layer must be at least as thick as the optical path difference divided by . The relaxation time of the liquid crystal orientation is approximately [12], where is the rotational viscosity and the splay elastic constant. Putting these together:
The relaxation time is proportional to the square of the scan span and the fourth power of the numerical aperture, and inversely proportional to the square of the usable index difference. The aperture of the liquid crystal lens and the relay magnification do not appear in the expression: reducing the aperture raises the power limit of a single-layer GRIN lens, but the required power increases in the same proportion, so the liquid crystal layer does not get any thinner.
Example. Take (assuming the entire birefringence is usable, which is an upper bound) and (example value). For NA 0.8 water immersion and a 20 μm span, the optical path difference is 4.8 μm, the liquid crystal layer about 24 μm and ms; for a 100 μm span, the optical path difference is 24 μm, the layer about 120 μm and s. If the effective NA is only 0.5 (water immersion), a 100 μm span needs an optical path difference of 9.4 μm and a layer of about 47 μm, with s; an NA 0.3 air objective with a 100 μm span needs 4.5 μm and about 22.5 μm, with ms.
These orders of magnitude agree with experiment: the liquid crystal lens of Martínez-Corral et al. used two 60 μm liquid crystal layers and had response times of 200–500 ms [5]. Lin et al. listed "the conflict between response time and focusing range" as the foremost problem for liquid crystal lens applications [10]. By comparison, layer switching with an elastic-membrane tunable lens takes about 15 ms [1].
GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously discusses several ways around this trade-off: split the liquid crystal into several stacked layers, which shortens the relaxation time of each by the square of the number of layers; let the phase reset at certain radii, or build a diffractive multi-ring-electrode lens, so that the liquid crystal layer only has to provide part of the optical path difference; or switch to a PB liquid crystal lens, whose layer is only micrometers thick but whose focal states are discrete. The reset and diffractive schemes are wavelength sensitive and are easier to use in a narrowband excitation path than in a broadband fluorescence path. In addition, the relaxation time applies to the voltage-off process; the voltage-on process can be accelerated by raising the drive voltage [12], and drive schemes are compared in Electrical Control of Liquid Crystal Devices: Response Time, Polarization Switches and Cascaded Switching.
Wavefront quality
Even setting aside the intrinsic spherical aberration of high-NA objectives, the phase profile of the liquid crystal lens itself is not automatically parabolic, and its residual higher-order terms add to the system aberrations (GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously). By the Maréchal criterion, the rms wavefront error budget of the whole system is about [6], to be shared among the liquid crystal lens, the relay and the objective; in confocal microscopy both excitation and detection pass through the liquid crystal lens, so its errors affect the point spread functions on both sides. For comparison, the electrowetting liquid lens used by Martínez-Corral et al. had a nominal maximum wavefront error of 50 nm [5], whereas their liquid crystal lens produced a markedly degraded point spread function [5].
The optical power of a liquid crystal lens is tunable, and its aberrations also change with voltage; a given set of voltages gives the best profile only near one focal shift. Each target focal shift therefore needs its own calibrated set of drive parameters, and the calibration must also account for the high-NA nonlinearity described above. Disclination lines that appear when voltage is applied scatter light, which in fluorescence imaging shows up as a raised background and reduced contrast.
Polarization
The polarization requirement depends on the beam path in which the liquid crystal lens is placed. When only excitation light passes through it (non-descanned detection in two-photon microscopy, reflection confocal microscopy), a single-layer liquid crystal lens aligned with the laser polarization is sufficient. When fluorescence passes through it (the shared path of fluorescence confocal microscopy, the detection path of widefield and light-sheet microscopy), a polarization-independent structure is required. An orthogonal dual-layer structure requires the two layers to be matched in thickness, drive and profile shape; otherwise the two polarization components have different focal shifts, which appears as polarization-dependent defocus. The total thickness, the number of interfaces and the number of drive channels also double.
Transmission
The photon budget of fluorescence imaging is tight. Loss of excitation light can be compensated by increasing laser power (within what the sample can tolerate), whereas loss of fluorescence directly lowers the signal-to-noise ratio. Each uncoated glass–air interface reflects about 4% at normal incidence (calculated from the Fresnel equations for an index of 1.5) [6].
Example. For an orthogonal dual-layer lens made of two separate liquid crystal cells with four uncoated outer surfaces, interface reflections alone reduce the transmission to about 85%, before counting absorption in the transparent electrodes and alignment layers.
The experience of Grewe et al. is instructive: in their tunable lens combination, the uncoated offset lens caused a reflection loss of about 9%, so they recommended coated lenses [1]. Index matching inside the liquid crystal lens, antireflection coating of the outer surfaces and absorption in the transparent electrodes should all be evaluated against measured transmission.
Aperture and field angle
The aperture of the liquid crystal lens is . As explained above, the focal shift range depends only on the optical path difference, not on the aperture; but the aperture affects performance through the field angle. The relay reduces the beam diameter and magnifies angles by the same factor, so the tilt of the chief ray of the field edge at the liquid crystal lens is approximately times its value at the objective's back focal plane.
Example. For a 40× objective ( mm) and a 500 μm diameter field of view, the edge chief ray is tilted by about at the back focal plane; with it is about at the liquid crystal lens, and with about .
The phase retardation of the liquid crystal layer depends on the angle of incidence; oblique incidence changes the local optical path difference and phase profile, which shows up as defocus or astigmatism at the edge of the field. In a scanning microscope, the scanned beam pivots precisely about the pupil conjugate plane, and the scan angle at the liquid crystal lens is magnified by in the same way. Conversely, choosing to reduce the angles requires the liquid crystal lens to have an aperture on the order of 8–17 mm (see in the table above), and in large-aperture GRIN liquid crystal lenses the field distribution in the center of the aperture is harder to control. A compromise between aperture and field angle is needed.
Temperature and calibration
The birefringence and viscosity of liquid crystals both vary with temperature [12], so the focal shift and the response time drift with temperature; long imaging sessions require temperature control or temperature-corrected calibration. Similar problems are common to tunable lenses: over 2000 back-and-forth switches, Grewe et al. observed the intensity of fluorescent beads fall to about 95%, which they attributed to a slight thermal drift of the tunable lens [1].
Timing of multi-plane imaging
The volume rate of multi-plane imaging is determined jointly by the acquisition time per layer and the switching time. The total sampling rate given by Grewe et al. is [1]
where is the number of layers, is the acquisition time of layer and is the switching time. With a switching time of 21 ms they reached 21.6 Hz over 40 positions in two layers, and 30 Hz when the switching time was shortened to 15 ms [1].
Example. For 4 layers at 33.3 ms each (30 frames per second): with a 1 ms switching time the volume rate is about 7.29 Hz and the acquisition duty cycle 97.1%; with 15 ms it is about 5.18 Hz with a duty cycle of 68.9%; with 300 ms it is only about 0.75 Hz with a duty cycle of 10%.
Liquid crystal lenses have another characteristic that differs from elastic-membrane lenses: the speeds in the two directions are asymmetric. When voltage is applied, a higher voltage can speed up molecular rotation; when it is removed, the only option is to wait for the elastic torque to pull the molecules back, which is usually much slower [12]. A sensible approach is to make every layer-to-layer switch in the voltage-on direction, so that each volume goes through only one slow relaxation, after the last layer, to return to the first layer, as in Figure 5.
Example. With the example parameters of Figure 5 (5 ms for a voltage-driven switch, a relaxation time constant of 40 ms and a wait of 3 time constants, i.e. 120 ms), the volume period for 4 layers is about 268 ms and the volume rate about 3.73 Hz. If every switch had to wait for a 120 ms relaxation, the volume period would become 613 ms and the rate only about 1.63 Hz.
If a layer switch can be completed within the scanner's flyback time, switching takes up almost no acquisition time. For applications that need frame-by-frame switching, the focal state switching of a PB lens (geometric phase lens) is set by a polarization switch or a very thin liquid crystal layer and is much faster than a thick GRIN liquid crystal layer, which makes it attractive in terms of timing; the cost is that the focal plane can take only a few predesigned values.
Comparison with other fast axial scanning methods
The table below compares liquid crystal lenses with several common fast axial scanning methods. Speeds and ranges are only specific results from the literature cited in this article and do not represent the limits of each technology.
| Method | Principle | Example speed or range in the literature | Main advantages | Main limitations |
|---|---|---|---|---|
| Piezo objective or Z stage | Moves the objective or the sample | About 10 Hz for 3D scanning with galvanometers [1] | No additional optical elements; image quality guaranteed by the objective | Speed limited by inertia; disturbs the sample and immersion liquid; vibration affects electrophysiological recording [1][2] |
| Elastic-membrane or electrowetting liquid lens | Electrically changes the curvature of a liquid surface | Elastic-membrane lens layer switching about 15 ms, range 0.7 mm [1]; tunable lens for light-sheet imaging at 30 volumes per second [3]; electrowetting lens response about 9 ms [5] | Polarization independent, large aperture, large range | Mechanical resonance and thermal drift [1]; focusing introduces defocus aberrations [7] |
| TAG acoustic gradient-index lens | An ultrasonic standing wave in a liquid creates a periodically varying index profile | Optical power varies continuously on the microsecond scale [13]; two-photon volume imaging at tens of Hz [14] | Extremely fast | The focus sweeps back and forth continuously and cannot stop at an arbitrary layer; requires phase-locked synchronization with the laser or detection [14] |
| Deformable mirror | Deformation of a reflective surface changes the wavefront | 20 kHz refresh rate, used for extended depth of field [15] | Reflective, no dispersion, polarization independent, can also correct aberrations | Requires a reflective beam path; the available optical path difference is limited by the mirror stroke |
| Remote focusing | A second objective replicates the sample space remotely, and a remote mirror is moved | Aberration-free refocusing over 70 μm at 1.4 NA [8]; 3D scanning at kHz rates over a range of several hundred micrometers [16] | No system spherical aberration at high NA | Requires a second objective and precise alignment; complex setup |
| Liquid crystal lens (GRIN type) | An electric field changes the liquid crystal index profile | Continuous 48 μm scan in confocal microscopy [11]; 200–500 ms response in widefield microscopy [5] | No moving parts, thin, low voltage, no mechanical resonance | Optical path difference and speed constrain each other; polarization dependent; image quality needs improvement [5][10] |
The table shows that liquid crystal lenses cannot compete with TAG lenses, deformable mirrors or remote focusing in speed, and currently also lag behind elastic-membrane tunable lenses in range and image quality. What sets them apart is that they have no moving parts or mechanical resonances, are very thin, need low drive voltages, and in principle have a phase profile that can be programmed with multiple electrodes. These features are most valuable when the required scan span is small or the objective NA is not high, when only narrowband, linearly polarized excitation light passes through the lens, when layer changes are infrequent or longer switching times are acceptable, and when the instrument needs a compact, vibration-free focusing module. Extended depth of field and focus stack acquisition are covered in Extending Depth of Field with Liquid Crystal Lenses: Focus Stacking and 3D Imaging; integration in compact instruments is covered in Liquid Crystal Lenses in Research Optics: Fiber Coupling, Beam Focusing and Wavefront Correction.
Demonstrated capabilities and potential yet to be verified
Demonstrated in the literature:
- A liquid crystal lens with a 0.65 NA objective achieved continuous axial scanning over 48 μm in a reflection confocal microscope and imaged micrometer-scale optical waveguide structures [11].
- A liquid crystal lens at a plane conjugate to the aperture stop of a widefield fluorescence microscope could move the object plane, but its image quality and 200–500 ms response time did not meet requirements at the time [5].
- The principle that a telecentric arrangement makes the focal shift linear in optical power while keeping the magnification constant has been verified with electrowetting liquid lenses and elastic-membrane tunable lenses [5][3]; the feasibility of tunable lenses for in vivo multi-plane two-photon imaging has also been demonstrated [1].
Application potential that still needs experimental verification:
- A single-layer liquid crystal lens in the excitation path of a two-photon microscope, exploiting the linear polarization and narrow bandwidth of the laser for multi-plane switching.
- A PB liquid crystal lens combined with a polarization switch, switching between a few focal planes on the sub-millisecond scale.
- Precompensation of high-NA spherical aberration, in addition to defocus, with multi-electrode liquid crystal phase devices.
Each of these directions can first be estimated from the relations in this article: compute the optical path difference from the scan span and the effective NA, compute the liquid crystal layer thickness and response time from the optical path difference and the material, and then determine the polarization and dispersion requirements from the position in the beam path.
References
[1] B. F. Grewe, F. F. Voigt, M. van 't Hoff, F. Helmchen. Fast two-layer two-photon imaging of neuronal cell populations using an electrically tunable lens. Biomedical Optics Express, 2(7): 2035–2046, 2011. DOI: 10.1364/BOE.2.002035
[2] E. J. Botcherby, R. Juškaitis, M. J. Booth, T. Wilson. An optical technique for remote focusing in microscopy. Optics Communications, 281(4): 880–887, 2008. DOI: 10.1016/j.optcom.2007.10.007
[3] F. O. Fahrbach, F. F. Voigt, B. Schmid, F. Helmchen, J. Huisken. Rapid 3D light-sheet microscopy with a tunable lens. Optics Express, 21(18): 21010–21026, 2013. DOI: 10.1364/OE.21.021010
[4] J. B. Pawley (ed.). Handbook of Biological Confocal Microscopy, 3rd ed. Springer, 2006. ISBN 978-0-387-25921-5. DOI: 10.1007/978-0-387-45524-2
[5] M. Martínez-Corral, P.-Y. Hsieh, A. Doblas, E. Sánchez-Ortiga, G. Saavedra, Y.-P. Huang. Fast axial-scanning widefield microscopy with constant magnification and resolution. Journal of Display Technology, 11(11): 913–920, 2015. DOI: 10.1109/JDT.2015.2404347
[6] M. Born, E. Wolf. Principles of Optics, 7th ed. Cambridge University Press, 1999. DOI: 10.1017/CBO9781139644181
[7] N. Koukourakis, M. Finkeldey, M. Stürmer, C. Leithold, N. C. Gerhardt, M. R. Hofmann, U. Wallrabe, J. W. Czarske, A. Fischer. Axial scanning in confocal microscopy employing adaptive lenses (CAL). Optics Express, 22(5): 6025–6039, 2014. DOI: 10.1364/OE.22.006025
[8] E. J. Botcherby, R. Juškaitis, M. J. Booth, T. Wilson. Aberration-free optical refocusing in high numerical aperture microscopy. Optics Letters, 32(14): 2007, 2007. DOI: 10.1364/OL.32.002007
[9] J. R. Lakowicz. Principles of Fluorescence Spectroscopy, 3rd ed. Springer, 2006. ISBN 978-0-387-31278-1. DOI: 10.1007/978-0-387-46312-4
[10] Y.-H. Lin, H.-S. Chen, M.-S. Chen. Electrically tunable liquid crystal lenses and applications. Molecular Crystals and Liquid Crystals, 596(1): 12–21, 2014. DOI: 10.1080/15421406.2014.918243
[11] S. A. Khan, N. A. Riza. Demonstration of a no-moving-parts axial scanning confocal microscope using liquid crystal optics. Optics Communications, 265(2): 461–467, 2006. DOI: 10.1016/j.optcom.2006.03.068
[12] D.-K. Yang, S.-T. Wu. Fundamentals of Liquid Crystal Devices, 2nd ed. John Wiley & Sons, 2014. ISBN 978-1-118-75200-5. DOI: 10.1002/9781118751992
[13] A. Mermillod-Blondin, E. McLeod, C. B. Arnold. High-speed varifocal imaging with a tunable acoustic gradient index of refraction lens. Optics Letters, 33(18): 2146–2148, 2008. DOI: 10.1364/OL.33.002146
[14] L. Kong, J. Tang, J. P. Little, Y. Yu, T. Lämmermann, C. P. Lin, R. N. Germain, M. Cui. Continuous volumetric imaging via an optical phase-locked ultrasound lens. Nature Methods, 12(8): 759–762, 2015. DOI: 10.1038/nmeth.3476
[15] W. J. Shain, N. A. Vickers, B. B. Goldberg, T. Bifano, J. Mertz. Extended depth-of-field microscopy with a high-speed deformable mirror. Optics Letters, 42(5): 995, 2017. DOI: 10.1364/OL.42.000995
[16] E. J. Botcherby, C. W. Smith, M. M. Kohl, D. Débarre, M. J. Booth, R. Juškaitis, O. Paulsen, T. Wilson. Aberration-free three-dimensional multiphoton imaging of neuronal activity at kHz rates. Proceedings of the National Academy of Sciences, 109(8): 2919–2924, 2012. DOI: 10.1073/pnas.1111662109