In research optics, liquid crystal lenses provide electrically tunable focus: a voltage changes the optical power and replaces the axial mechanical adjustment used in fiber coupling, beam focusing and defocus correction. Much of the "focusing" on an optical table is done by hand: turning the axial micrometer of a fiber coupling mount until the power meter reading peaks; pushing a collimating lens back and forth until the beam has the right size on a detector several meters away; moving a sample stage until a spectral peak in the spectrometer is strongest. Each of these is a mechanical degree of freedom that is locked once adjusted. When the ambient temperature changes, or the light source or sample is replaced, it has to be adjusted again; to automate the experiment, these adjustments have to be converted into motorized translation stages.
A liquid crystal tunable lens (LC-ETL) can replace the "movement along the optical axis" class of these degrees of freedom with a voltage: a thin element with electrically controlled optical power placed in a collimated or nearly collimated beam changes the vergence of the beam, and the downstream focus or beam waist moves back and forth accordingly. This article examines how far this replacement can go in research optics, optical metrology and compact analytical instruments. It first reviews the typical focusing tasks and expands on several of them: beam waist positioning, fiber coupling, defocus correction in wavefront sensing and sampling in portable spectroscopy. It then discusses the programmability and closed-loop control that electrical control brings, and how response time, hysteresis, temperature drift and calibration limit the closed loop. Next it explains how liquid crystal lenses, liquid crystal spatial light modulators and deformable mirrors divide the work of wavefront correction. Finally it discusses modular integration and gives a checklist for selection and integration.
For the structural types and main challenges of liquid crystal lenses, see the review [1]. Topics already covered in this series are only cited here by their conclusions: the difference between focusing and zooming, and the relationship between the optical power limit and the aperture, in What Is a Liquid Crystal Tunable Lens? Principles, Types and System Integration; the principle of continuous focusing in GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously, and discrete focal states in PB Lens Principles: How Geometric Phase Liquid Crystal Lenses Switch Focal Length Electrically; the -parameter analysis of Gaussian beam focus shift, confocal displacement sensing and the limits under high-power lasers in Dynamic Laser Focusing with Liquid Crystal Lenses: Focus Control in Processing and Inspection.
Conventions used in this article: the optical power is , in diopters D (i.e. ), with converging power positive; the optical power added by the liquid crystal lens is written . The beam radius is the radius at which the intensity falls to of its central value, the waist radius is , and the Rayleigh length is ; all are calculated for an ideal fundamental-mode Gaussian beam. The relaxation time of the liquid crystal layer is estimated as , where is the liquid crystal layer thickness, and the example viscoelastic coefficient used throughout this series is adopted; the derivation is in Electrical Control of Liquid Crystal Devices: Response Time, Polarization Switches and Cascaded Switching.
Focusing tasks in laboratory optics
The quantities that need adjusting along the optical axis in research optics can be grouped by "what is adjusted and what is observed" as follows.
| Task | Adjusted quantity | Common feedback signal | Role of the liquid crystal lens |
|---|---|---|---|
| Beam collimation and waist positioning | Position and size of the beam waist | Spot size measured by a camera or knife edge | Changes the incident vergence |
| Waist scanning and beam diagnostics | Position of the focus along the optical axis | Spot radius at each position | Replaces moving the camera or the lens |
| Fiber coupling | Axial distance between the focus and the fiber end face | Coupled power | Compensates axial mismatch |
| Spectroscopic and Raman sampling | Focus on the sample, image on the slit | Signal or spectral peak intensity | Follows the sampling distance |
| Focus tracking in OCT and ranging | Depth of the focus | Signal intensity | Makes the focus follow the imaging depth |
| Wavefront sensing and adaptive optics | Defocus component of the wavefront | Defocus coefficient from the wavefront sensor | Takes care of defocus and a few low-order aberrations |
These tasks have three things in common. First, what is adjusted is mostly the vergence of a collimated or nearly collimated beam; the beams are narrow, the intensity is moderate, and the source is often a linearly polarized narrowband laser. This falls squarely within the most favorable conditions for liquid crystal lenses: the optical power limit is high at small apertures, a single-layer device aligned with the laser polarization loses no light, and a narrowband source also minimizes the cost of chromatic aberration. Second, the required optical power is usually very small; as shown below, moving a beam waist over a long optical path needs only a few tenths of a diopter, and compensating the axial mismatch in fiber coupling is also within 1 D. Third, most of them already have a feedback signal and can be run in closed loop. Figure 1 shows the four types of optical path discussed below.
Beam collimation and waist positioning
The waist of a weak lens is not at the geometric focus
The waist of a collimated laser beam is usually near the collimating lens. Experiments often require placing the waist somewhere downstream, for example at the center of a nonlinear crystal, on a detector several meters away, or at a position required by a resonator. The geometric-optics intuition is that adding a lens with optical power focuses the light at . For a Gaussian beam, this holds only when is large enough.
Describe the Gaussian beam with the complex parameter : , where is the radius of curvature of the wavefront; at the waist , and at distance from the waist . A thin lens subtracts its optical power from [2]. Suppose the waist of the incident beam lies exactly at the liquid crystal lens; after the lens
Take the reciprocal and write it in the form , where is the distance from the new waist to the liquid crystal lens and is the new Rayleigh length. The real part gives the waist position and the imaginary part the new Rayleigh length; since the square of the waist radius is proportional to the Rayleigh length,
The Gaussian beam lens imaging formula given by Self is equivalent to this, and Self also shows how it corresponds to the geometric-optics imaging formula [3].
Only one dimensionless quantity appears in the expression, , the ratio of the Rayleigh length of the incident beam to the focal length of the lens. It determines which regime the result falls into:
- : , back to geometric optics, with the waist near the focal point of the lens.
- : , the waist position is proportional to the optical power and much shorter than the geometric focal length .
- : reaches its maximum value , and the waist radius shrinks to of its original value.
This maximum means that a positive lens placed at the waist, however weak, cannot push the waist beyond . The reason is diffraction. Without a lens, the beam already spreads over distances of order ; when the focal length of the lens is much longer than , the convergence it introduces is weak, and diffraction-induced spreading takes over before the beam can contract to the geometric focus, so the narrowest point appears much closer than the geometric focus.
Example. For a 633 nm beam with mm, m and D. With 0.05 D added, the geometric focus is at 20 m, but the actual waist is at about 1.16 m, with a radius about 0.97 of the original; with 0.2 D added, the waist is at about 2.48 m, close to the farthest reachable position, with a radius about 0.71 of the original; with 1 D added, the waist is at about 0.96 m, already close to the geometric value of 1 m, with a radius about 0.20 of the original.
Figure 2(a) shows the transition between the two regimes: on the left the curves rise with slope (), on the right they approach the geometric asymptote with slope , and the peak lies at .
This has two implications for experimental design. First, the optical power needed to adjust the waist position over a long optical path is very small, and so are the demands on the liquid crystal layer. The whole rising branch, , needs an optical path difference of no more than at the waist radius ( equals exactly at ), about 101 nm at 633 nm. This is the same fact as the statement in Dynamic Laser Focusing with Liquid Crystal Lenses: Focus Control in Processing and Inspection that "moving the focus by one Rayleigh length takes only about of optical path difference". With such a small phase, the liquid crystal layer can be very thin, and the response is correspondingly fast. Second, placing the waist beyond requires first expanding the beam to increase , or adding another lens downstream; adjusting the optical power of a single lens alone cannot do it. The most suitable role for the liquid crystal lens here is therefore this: fixed beam-expanding and collimating optics set the approximate waist position, and the liquid crystal lens provides fine adjustment and drift compensation on top of that.
Mode matching needs two degrees of freedom
A Gaussian beam is determined by two real parameters, for example the size and position of the waist. Matching a beam to the eigenmode of a resonator or to the mode field of another fiber requires both parameters to match at the same time [2]. The expression above shows that changing the of a single lens changes and together; they cannot be set separately. Setting both independently requires at least two tunable lenses some distance apart. This is the same reason that electrically controlled zooming needs at least two tunable lenses. In practice, the main mode matching is more commonly done by a fixed lens group, and the liquid crystal lens only compensates small mismatches due to temperature drift or a change of light source; such mismatches show up mainly as a shift of the waist position, and one liquid crystal lens is enough.
Waist scanning and beam quality measurement
Measuring the waist, divergence and beam quality factor of a beam requires measuring the spot radius at several positions on both sides of the waist and fitting the results with the Gaussian beam propagation law. The common practice is to move a camera along the optical axis on a translation stage, or to move the focusing lens. Dynamic Laser Focusing with Liquid Crystal Lenses: Focus Control in Processing and Inspection showed with the parameter that when the liquid crystal lens is at the front focal plane of the focusing lens, adding only translates the waist as a whole by , with the waist radius and Rayleigh length unchanged. This is exactly the condition a measurement needs: keeping the camera still and scanning the voltage is equivalent to moving the camera along the optical axis.
Example. A collimated 1064 nm beam with a radius of 1.5 mm and a focusing lens with a focal length of 100 mm give mm at the focus and a waist radius of about 22.6 μm. Covering 3 Rayleigh lengths on each side of the waist (about mm) requires about D; the optical path difference at the beam radius is about 508 nm, and about 1.14 μm if the clear radius is taken as 1.5 times the beam radius. With a usable index difference , the liquid crystal layer is about 7.6 μm thick, with a relaxation time of about 24 ms.
This use currently belongs to application potential; in the literature searched for this article, no report of waist-scanning measurements with a liquid crystal lens was found. Two points need attention: the deviation of the liquid crystal lens's own phase from a parabola is superimposed on the beam being measured, and its fourth-order phase aberration makes the measured too large; and the relationship between voltage and optical power must be calibrated in advance with a known beam.
Fiber coupling: handing the axial position to a voltage
Tolerance to axial mismatch
When free-space light is coupled into a single-mode fiber, the coupling efficiency equals the normalized overlap integral of the incident field and the fiber mode field [4]. Approximating both as fundamental-mode Gaussian beams, with radii and and wavefront radii of curvature and in the same plane, the overlap integral can be written as
The first term in the denominator comes from spot size mismatch and the second from wavefront curvature mismatch. When the waist of the focused spot equals the fiber mode field and is only offset by along the optical axis, then at the fiber end face one beam is at its waist and the other is away from its waist; substituting the Gaussian beam and , the two terms together simplify to
where is the Rayleigh length in air of a fiber mode field of radius . For a lateral offset , the same integral gives . A liquid crystal lens can compensate only axial mismatch; lateral and angular misalignment still require mechanical adjustment or a separate beam steering element.
Example. At 1550 nm with a mode field radius of 5.2 μm (an example value, of the order typical of single-mode fibers at this wavelength), μm. requires μm, and requires μm.
Converting to optical power
With a coupling lens of focal length and the liquid crystal lens placed in the collimated path, adding moves the focus by about , independent of the position of the liquid crystal lens to first order. Substituting into the expression above, the coupling efficiency as a function of the added optical power is
is the added optical power that places the focus exactly on the fiber end face. If the input light is not strictly collimated but has a vergence (positive for converging, negative for diverging), it changes only the real part of , just like a thin lens, and adds directly to ; the optimum then becomes . One approximation is used here: when the vergence is small, the spot size at the focus is essentially unchanged; this holds exactly if the liquid crystal lens is at the front focal plane of the coupling lens.
Example. Take a coupling lens with a focal length of 11 mm (example value). The collimated beam radius that matches the fiber above is mm, and . corresponds to D, and to 0.91 D. If the same coupler is switched to a source with a vergence of D (example value), the focus moves back by about 60.5 μm without adjustment and the efficiency drops to about 77%; adding D with the liquid crystal lens restores it, see Figure 2(b).
With such a narrow beam, the burden on the liquid crystal lens is light. Taking the clear radius as 1.5 times the beam radius, about 1.57 mm, 0.5 D needs an optical path difference of only about 0.61 μm at the edge; with , the liquid crystal layer is about 4.1 μm thick, with a relaxation time of about 7 ms.
Polarization: a limitation that can also be exploited
A single-layer GRIN gradient-index liquid crystal lens focuses only the extraordinary wave polarized along the director; the orthogonal ordinary wave sees a uniform flat plate. With a polarization-maintaining source whose polarization is aligned with the director, no light is lost; with a randomly polarized source, the focus of the ordinary component does not move, and the coupling efficiencies of the two components have to be calculated separately.
Chen, Chen and Lin made use of exactly this [5]. They sandwiched a liquid crystal lens between two GRIN rod lenses to make a coupler: with no voltage, light is focused into the single-mode fiber; with a voltage applied, the optical power for the extraordinary wave decreases, the spot grows, and the coupling efficiency falls continuously, so the device becomes a variable optical attenuator for this polarization component; the optical power for the ordinary wave is unchanged, and its coupling efficiency stays high. For input light of arbitrary polarization, the device therefore also becomes a broadband polarizer, with a polarization-dependent loss above 30 dB and a switching time of about 1 s [5]. This work shows that a liquid crystal lens can electrically change the coupling state of a single-mode fiber; conversely, if the goal is to maintain maximum coupling, the input polarization must be aligned with the director, or a double-layer structure with orthogonal alignment must be used.
Closed loop with coupled power as feedback
The coupling efficiency is flat near its peak, , so a single power reading cannot tell which direction to adjust. Two methods are common.
- Scan search. Change in steps, record the power, and fit a few points near the peak to find its position. This is the same as the hill-climbing search in camera contrast autofocus; see Liquid Crystal Lens Autofocus: Electrical Focusing in Cameras and Machine Vision.
- Dither locking. Superimpose a small periodic modulation on the operating point and use lock-in detection to extract the component of the power at frequency . From the expression above, this component is proportional to : it is zero at the peak, its sign indicates the direction of adjustment, and its amplitude is proportional to the deviation, so it can be used directly as an error signal.
Example. With the of the example above and a dither amplitude of 0.05 D, the instantaneous efficiency loss at the two extremes of the dither is about 0.3%.
The dither frequency must be much lower than the response speed of the liquid crystal lens; otherwise the optical power cannot follow the modulation, and the locking bandwidth is only a small fraction of the liquid crystal response speed. For the thin device of about 7 ms in the example above, several corrections per second are possible; for thick devices with relaxation times of several hundred milliseconds or more, the closed loop is suitable only for compensating slow changes on the scale of minutes, such as temperature drift or a change of source, not for suppressing vibration.
Electrical control: from manual degrees of freedom to programmable ones
What it brings
Replacing an axial adjustment with a voltage changes four things.
- Programmability. The voltage can be set, logged and replayed by a script. The focus position needed at each step of an experimental procedure can be written into a recipe, and waist scans and focus scans become a loop in a program.
- Closed-loop operation. Given a feedback signal, the system can find and lock onto the peak automatically, or measure and then correct directly; Figure 3 summarizes the components of the closed loop.
- No mechanical backlash or wear. Micrometers and lead screws have backlash, reading errors and wear; a liquid crystal lens has no moving parts, and long-term repeated switching does not wear in any play. But it has no "position sensor" either: the optical power at a given voltage depends on temperature and device state, so repeatability has to be ensured by calibration and closed-loop control rather than by mechanical structure.
- Simpler structure. Without translation stages, micrometers, motors and motion controllers, the optical path can be shorter, the components can be fixed in a lens tube, and vibration resistance improves. Electrically, the liquid crystal layer is approximately a small capacitor, and the drive power is very low; see the estimate in What Is a Liquid Crystal Tunable Lens? Principles, Types and System Integration.
What limits the closed loop
Response time. The relaxation time of the liquid crystal orientation is proportional to the square of the liquid crystal layer thickness; the switch-on direction can be sped up by raising the voltage or applying a short overdrive pulse, but the switch-off direction can only wait for elastic recovery [6]. Each step of the closed loop must wait for the liquid crystal to settle before reading the sensor; otherwise it reads a transitional state. This is especially pronounced in thick modal liquid crystal lenses: the modal liquid crystal lens of Simonov et al. had a 40 μm liquid crystal layer and took about 16 s to settle completely under normal driving, mainly because of a transient overshoot of the aberration coefficients; switching the drive frequency shortened this to about 4 s [7]. When designing a closed loop, therefore, first determine the liquid crystal layer thickness from the required optical path difference, then estimate the waiting time for each step from the thickness, and only then set the control bandwidth.
Hysteresis. The electrically controlled birefringence structure of a nematic liquid crystal has no intrinsic bistability; its steady-state orientation is determined by the rms drive voltage, the frequency and the temperature, so in principle there is no backlash in the mechanical sense. Two kinds of effects can nevertheless make the result depend on history: when the drive voltage contains a DC component, ions in the liquid crystal accumulate at the alignment layer interfaces and screen the applied field, causing image sticking and slow drift [6]; and in hole-patterned electrode structures, disclination lines may appear while the voltage is applied, making the phase distribution deviate from the intended one; see GRIN Liquid Crystal Lenses: How Voltage Tunes Focal Length Continuously. Calibration should therefore measure the voltage–optical power curve separately in the increasing and decreasing directions, to confirm whether the steady states at the same voltage agree.
Temperature drift. The birefringence of a liquid crystal decreases as temperature rises. In Haller's approximation, the relative temperature coefficient of the birefringence is , where is the clearing point and a material constant [8]. Under fixed driving, the optical power of a GRIN lens changes roughly in the same proportion as the birefringence.
Example. Taking and K (example values), the relative temperature coefficient at 25 °C is about . For a liquid crystal lens operating at 0.5 D, a 5 K temperature rise reduces the optical power by about 0.010 D. For the fiber coupling above (), the efficiency loss is only about 0.01% and negligible. The same 0.010 D error on a pupil of radius 6 mm gives an rms defocus wavefront of about 52.5 nm (see the next section for the conversion), already above the nm criterion at 633 nm; by the Maréchal criterion, when the rms wavefront error does not exceed , the Strehl ratio is about 0.8 and imaging is close to the diffraction limit [9].
The same temperature drift can be negligible in one task and unacceptable in another. What decides it is the task's tolerance on : for fiber coupling the tolerance is set by , and for wavefront correction by the allowed rms defocus and the pupil radius.
Calibration. For open-loop use, the relationship between optical power and voltage (and also frequency for modal devices) must be measured for each device, at several temperatures, and stored as a lookup table that is interpolated by temperature during operation. The optical power can be measured directly as the defocus coefficient with a Shack–Hartmann sensor, or converted from the measured displacement of the focus. For closed-loop use, the lookup table is still useful: it provides a starting point for the search and shortens the time to find the peak.
Defocus and wavefront correction: how liquid crystal lenses, spatial light modulators and deformable mirrors divide the work
Defocus is the fourth Zernike term
Wavefront aberrations are commonly expanded in Zernike polynomials. With Noll's numbering and normalization, the defocus term is [10]
where is the normalized pupil radius and is the pupil radius. Noll's normalization gives each term an rms value of 1 over the unit circle, so the expansion coefficient equals the rms wavefront error contributed by that term.
A thin lens of optical power adds an optical path of over the pupil (the sign depends on the wavefront sign convention; only the magnitude matters here). Writing as , the parabolic phase decomposes into a constant (the piston term, which does not affect imaging) and a pure defocus term, with the defocus coefficient
The peak-to-valley value is . An ideal liquid crystal lens therefore provides pure Z4, without any other term.
Example. For a pupil radius of 5 mm and D, μm, about 0.57 waves at 633 nm, with a peak-to-valley value of 1.25 μm.
This conversion can be used to check data in the literature. The modal liquid crystal lens of Simonov et al. had a clear aperture of 5 mm and was measured with a Shack–Hartmann sensor with 127 subapertures; the paper reports a tunable defocus of about 4 waves over a 4.9 mm aperture at 534 nm, corresponding to about 2.51 D [7]. By the expression above, 2.51 D over a radius of 2.45 mm corresponds to μm, about 4.07 waves, consistent with the value in the paper.
Which modes a liquid crystal lens can provide
A liquid crystal lens with a hole-patterned electrode or modal control is rotationally symmetric, and its phase distribution expanded in the radius contains only even-order terms such as , and , corresponding to rotationally symmetric modes such as Z4, Z11 (primary spherical aberration) and Z22 (secondary spherical aberration); see Figure 4(a). The part of the actual phase distribution that deviates from a parabola appears mainly as spherical aberration.
This is a defect, but it can also become a usable degree of freedom. The modal control proposed by Naumov et al. uses a high-resistivity layer and an AC voltage applied only at the boundary of the aperture to produce a distributed voltage, and both the amplitude and the frequency of the voltage can change the shape of the phase distribution [11]. Simonov et al. used this to tune spherical aberration together with defocus: the measured wavefront consisted mainly of defocus and spherical aberration, and the spherical aberration coefficient could be tuned between and waves [7]. Naumov et al. also made a cylindrical modal lens [11]; two cylindrical lenses with orthogonal axes could in principle be combined to produce astigmatism, which belongs to application potential. Non-rotationally-symmetric modes such as coma and trefoil cannot be produced by a rotationally symmetric liquid crystal lens.
Division of work with liquid crystal spatial light modulators and deformable mirrors
Liquid crystal spatial light modulators. They write phase point by point on a pixel array and can produce arbitrary phase distributions. Love used a nematic liquid crystal spatial light modulator to demonstrate open-loop wavefront correction, and produced and quantitatively measured the first 15 principal Zernike terms [12]. When it produces defocus, the phase wraps at and becomes a Fresnel-type diffractive lens. The local period over which the phase changes by is , smallest at the pupil edge; each period needs at least pixels to maintain diffraction efficiency, and with a pixel pitch the maximum optical power that can be written is
The first-order diffraction efficiency of an -level staircase approximation is [13].
Example. For a pixel pitch of 8 μm, a pupil radius of 6 mm and a wavelength of 633 nm: with (the sampling limit), the limit is about 6.6 D, but the efficiency is only about 41%; with , it is about 1.65 D, with an efficiency of about 95%. By comparison, a GRIN liquid crystal lens with the same 12 mm aperture and an optical path difference of 6 μm has an optical power limit of only about 0.33 D.
In monochromatic light, then, the defocus a spatial light modulator can write is not small. Its costs lie elsewhere: the wrapped phase is exactly only at the design wavelength, and the optical power is proportional to wavelength, a diffractive dispersion like that of a PB lens; liquid crystal on silicon devices are reflective and need relay optics to place them at a plane conjugate to the pupil; the pixelated phase produces higher diffraction orders; and it too acts on only one linear polarization direction. A GRIN liquid crystal lens, by contrast, is transmissive and can be inserted directly into a collimated path; its phase is continuous, without wrapping or pixel diffraction, its dispersion comes only from the dispersion of the liquid crystal's refractive index itself, and it needs only one or two drive voltages.
Deformable mirrors. They change the shape of a reflective surface with actuators, are polarization-independent, work over a very wide spectral range, and can correct many orders of aberration simultaneously [14]. Their defocus capability is limited by actuator stroke; because reflection doubles the optical path, the mirror deformation only needs to be half of the wavefront peak-to-valley value.
Example. For a pupil radius of 6 mm and a defocus correction of 0.5 D, the wavefront peak-to-valley value is μm and the mirror deformation peak-to-valley value is 4.5 μm, which should be compared with the stroke of the chosen deformable mirror. If the defocus is taken over by a liquid crystal lens, the stroke of the deformable mirror can be reserved for other aberrations.
The composition of the aberration determines the division of work. Noll calculated the residual phase variance of a wavefront distorted by atmospheric turbulence (Kolmogorov spectrum) after correcting successive terms [10]. Figure 4(b) shows that after removing the two tilt terms the residual is , and removing defocus as well only lowers it to , a reduction of about 17%; even after correcting up to the 11th term, remains. Here is the aperture, the atmospheric coherence length, and the variance is in . For this kind of random aberration, defocus is only one of many terms, and a liquid crystal lens alone cannot form an adaptive optics system. Another class of aberration common in research instruments is dominated by defocus: changes in sample depth, temperature-induced focus drift and focus shifts after a wavelength change all appear first as Z4. In this case the liquid crystal lens can be the main correction element, or it can work with a deformable mirror or spatial light modulator and take care of defocus alone, leaving the stroke or phase range of the latter for higher-order aberrations.
| Comparison | GRIN or modal liquid crystal lens | Liquid crystal spatial light modulator | Deformable mirror |
|---|---|---|---|
| Operation | Transmissive, continuous phase | Reflective for silicon-based devices, pixelated phase, wrapped at | Reflective, surface deformation |
| Available modes | Defocus, plus spherical aberration under modal control | Arbitrary distributions, limited by pixel count | Multiple orders, limited by actuator count and stroke |
| Polarization | A single layer acts on only one linear polarization component | Acts on only one linear polarization component | Independent |
| Dispersion | From the dispersion of the liquid crystal refractive index | Diffractive dispersion from phase wrapping | Reflective, essentially dispersion-free |
| Integration | Inserted directly into a collimated path | Needs relay to a plane conjugate to the pupil | Needs relay, folds the optical path |
| Main limitation | Optical power inversely proportional to the square of the aperture; thick layers respond slowly | Efficiency falls with fewer pixels per period; higher diffraction orders | Stroke and actuator count |
Measure, then correct
With a wavefront sensor, a defocus closed loop needs no search. A Shack–Hartmann sensor divides the pupil into many subapertures with a lenslet array, measures the local wavefront slope in each subaperture, and reconstructs the Zernike coefficients from them [14]. Once is measured, the magnitude of the required added optical power is , with the sign opposite to the measured defocus; this is the structure in Figure 1(c). Because liquid crystal lenses show transient overshoot [7], practical control usually corrects only part of the measured error at each step, converging gradually and waiting for the liquid crystal to settle at each step.
Compact analytical instruments: what structural simplification means
Portable spectroscopy and Raman probes
The Raman signal is weak, and the collection efficiency depends directly on whether the excitation focus lies on the sample. Yang et al. pointed out that existing fiber-optic Raman probes are mostly fixed-focus non-contact probes, which must hold the probe-to-sample distance precisely to collect enough signal; they integrated a liquid lens autofocus unit into a hand-held fiber-optic Raman probe, using a distance sensor and an algorithm they developed to determine the probe-to-sample distance, and markedly improved the signal stability during hand-held operation [15]. This shows that compact spectroscopic instruments do need electrically tunable focusing. Achieving the same function with a liquid crystal lens currently belongs to application potential, and the following conditions need to be considered.
- Sampling geometry. The excitation focus and the collection focus must move together, so the liquid crystal lens should be placed in the path shared by excitation and collection, as in Figure 1(d); if it is placed only in the excitation path, the conjugate point of the collected light does not move, and the signal actually decreases.
- Polarization. The excitation light is a linearly polarized laser, and a single-layer liquid crystal lens only needs to be aligned with its polarization. But Raman scattered light is generally only partially polarized, and a single-layer liquid crystal lens in the shared path does not act on its ordinary component, whose collection focus therefore does not move. This is the same situation as in fluorescence confocal microscopy, and an orthogonal double-layer structure is needed; see Microscope Z-Scanning with Liquid Crystal Lenses: Fast Focusing Without Moving the Objective.
- Spectral range. Raman scattered light covers a fairly wide range of wavelengths.
Example. With 785 nm excitation, Stokes Raman shifts of 200–2000 correspond to about 797.5–931.2 nm. The optical power of a PB lens is proportional to wavelength and differs by about 17% between the two ends of this range, and by up to about 19% relative to the excitation wavelength, so the excitation focus and the collection focus would be clearly separated. The optical power of a GRIN liquid crystal lens changes only with the dispersion of the liquid crystal birefringence, which must be checked for the specific material, but it has no change proportional to wavelength of the kind that diffractive devices have. Chromatic aberration compensation over a broad band is discussed in Applications of Achromatic Liquid Crystal Devices: Full-Color Displays and Broadband Polarization Control.
Focusing at the spectrometer entrance is the same problem: the spot on the sample has to be imaged onto the entrance slit, and when the image is defocused it widens at the slit and the part beyond the slit is blocked. For fiber-coupled spectrometers, the problem reverts to the fiber coupling of the previous section, except that the tolerance of multimode fibers is much wider.
Focus tracking in OCT and ranging
Optical coherence tomography (OCT) selects depth with a coherence gate, and its axial imaging range can reach the millimeter scale; its lateral resolution, however, is set by the focused spot and is maintained only within the depth of focus. The smaller the focus, the shorter the depth of focus.
Example. For a spot diameter of 3 μm (waist radius 1.5 μm), a wavelength of 800 nm and a tissue refractive index of 1.4 (all example values), the depth of focus is μm; maintaining this lateral resolution over a depth of 2 mm requires about 81 focal plane positions.
Murali et al. first demonstrated experimentally the benefit of refocusing at multiple depths: in highly scattering biological samples, optical coherence microscopy maintained an axial resolution of 2.5 μm and a lateral resolution of 6.5 μm throughout the sample [16]. They then presented an optical design for a hand-held probe with an embedded addressable liquid crystal lens and no moving parts, which can focus to a depth of 2 mm in skin-like tissue and maintain a resolution of 3 μm over an imaging volume of 8 mm³; the light source is a Ti:sapphire pulsed laser with a center wavelength of 800 nm and a bandwidth of 100 nm, and the design also compensates on-axis and off-axis aberrations at different depths [16]. Note that the probe in that paper is an optical design result. A broadband source means the dispersion of the liquid crystal lens must be included in the design, consistent with the discussion of Raman collection light above. Fusion of images from multiple focal planes is covered in Extending Depth of Field with Liquid Crystal Lenses: Focus Stacking and 3D Imaging.
Code reading and inspection modules
Barcode and character reading modules need fast focusing between different reading distances, as discussed in Liquid Crystal Lens Autofocus: Electrical Focusing in Cameras and Machine Vision. Bioassay modules such as microfluidic chip readers and fluorescence plate readers face a similar problem: thickness tolerances and placement errors of consumables make the position of the detection surface slightly different each time, and the system must focus on the channel or the well bottom. Such modules have short optical paths and small stops, which suits liquid crystal lenses; but fluorescence is unpolarized, so a liquid crystal lens in the detection path needs a polarization-independent structure. All of these uses currently belong to application potential.
Where the structural simplification shows
In compact instruments, the value of a liquid crystal lens comes mainly from "what is removed": there is no longer any need to reserve travel, guides and actuators for focusing, no position encoder is needed, and the optical module can be made as a sealed, fixed unit; with no moving mass, it is insensitive to shocks and changes of orientation during hand-held use; and the drive is a low-power capacitive load, suitable for battery operation. The price is the need for drive electronics, temperature measurement and calibration data, plus the polarization conditions discussed above.
Modularity: integrating into standard optical platforms
Research optics are mostly built on standard lens tube or cage systems. Making a liquid crystal lens into a plug-in module requires handling the interfaces properly in four respects: optical, mechanical, electrical and control.
Optical position. In a collimated path, the focus shift caused by the added optical power is independent of the position to first order; at the front focal plane of the focusing lens, the focus shift is strictly linear and the spot size unchanged. In imaging or relay paths, the lens should be placed at a plane conjugate to the pupil to keep the magnification unchanged. In every case it is best placed where the beam is narrow to reduce the required aperture, but not near a focus, to avoid excessive power density.
Mechanical mounting. A liquid crystal lens is a flat element that fits into a standard lens tube or a cage-system mounting plate. A single-layer device is sensitive to the polarization direction, so the mount should be rotatable about the optical axis, and the director direction should be marked on the housing for alignment with the laser polarization. To keep back-reflections from returning to the laser or fiber, the device can be tilted slightly; but oblique incidence changes the effective refractive index and introduces astigmatism, so the tilt should be as small as possible and the device calibrated at the actual tilt.
Electrical interface. The driver outputs an AC voltage without a DC component, with adjustable amplitude; modal devices also need an adjustable frequency [6]. A device with a single hole-patterned electrode needs only two wires, a device driven by two voltages needs three, and an orthogonal double-layer device doubles that if the layers are driven separately. The temperature sensor should be in close contact with the liquid crystal cell.
Control interface. It is more convenient for users if the driver accepts "added optical power" rather than "voltage": internally it converts by temperature-dependent lookup, and after the liquid crystal has settled it issues a ready signal for synchronizing data acquisition. Higher-level experiment scripts or instrument software then deal only with optical quantities, and no program changes are needed after a device is replaced or recalibrated.
Selection and integration checklist
| Item | What to check | Basis or conversion |
|---|---|---|
| Aperture | Whether the clear aperture covers the beam; at least 1.5 times the beam radius is recommended | The optical power limit is inversely proportional to the square of the aperture |
| Optical power range | The range the task needs, and whether both positive and negative are needed | Of order for waist positioning; converted via for fiber coupling |
| Optical path difference | Whether the edge optical path difference is within the device's capability | times the square of the clear radius, divided by 2 |
| Polarization | Polarization state of the source, whether it can be aligned with the director, whether an orthogonal double layer is needed | A single layer acts on only one linear polarization component |
| Wavelength | Optical power and transmittance at the operating wavelength and bandwidth | The optical power of diffractive and PB types is proportional to wavelength |
| Power | Average power, peak power density, electrode absorption and thermal lensing | Estimate from damage threshold and absorption heating |
| Response time | Whether the settling time of each step meets the cycle time and closed-loop bandwidth | Relaxation time is proportional to the square of the liquid crystal layer thickness |
| Drive | Number of channels, voltage, whether frequency tuning is needed, absence of DC component | Modal devices use two parameters, amplitude and frequency |
| Wavefront quality | Residual aberration in each drive state | Maréchal criterion |
| Temperature | Operating temperature range, temperature compensation in open loop | Relative temperature coefficient of the birefringence |
| Calibration | Lookup tables for increasing and decreasing voltage at several temperatures | Still used as a starting point in closed loop |
Demonstrated capabilities and application potential
Verified in the literature or presented as designs:
- A fiber coupler made of a liquid crystal lens and GRIN rod lenses can electrically change the coupling efficiency of a single-mode fiber and serve as a variable optical attenuator and broadband polarizer, with a switching time of about 1 s [5].
- A modal liquid crystal lens provides about 2.5 D of tunable defocus over a 5 mm aperture while also tuning spherical aberration, measured with a Shack–Hartmann sensor, with a settling time of about 4 s [7].
- A nematic liquid crystal spatial light modulator can produce the first 15 principal Zernike terms and perform open-loop wavefront correction [12].
- An optical design of a hand-held optical coherence microscopy probe with an embedded liquid crystal lens and no moving parts [16].
Still application potential that requires experimental verification:
- Using a liquid crystal lens to finely adjust the waist position over long optical paths, to compensate drift in mode matching, and to perform waist-scanning measurements of .
- Maintaining fiber coupling efficiency by dither locking with the coupled power as feedback.
- Having a liquid crystal lens take care of defocus alone in an adaptive optics system, saving stroke or phase range for a deformable mirror or spatial light modulator.
- Replacing liquid lenses or mechanical focusing with liquid crystal lenses in portable Raman probes and bioassay modules.
All of these directions can first be estimated with the relationships in this article: calculate the required and optical path difference from the task's tolerance, the liquid crystal layer thickness and response time from the optical path difference and the material, then determine the polarization and dispersion requirements from the position in the optical path, and finally decide between open-loop lookup and closed-loop control.
References
[1] Y.-H. Lin, Y.-J. Wang, V. Reshetnyak. Liquid crystal lenses with tunable focal length. Liquid Crystals Reviews, 5(2): 111–143, 2017. DOI: 10.1080/21680396.2018.1440256
[2] H. Kogelnik, T. Li. Laser beams and resonators. Applied Optics, 5(10): 1550, 1966. DOI: 10.1364/AO.5.001550
[3] S. A. Self. Focusing of spherical Gaussian beams. Applied Optics, 22(5): 658, 1983. DOI: 10.1364/AO.22.000658
[4] R. E. Wagner, W. J. Tomlinson. Coupling efficiency of optics in single-mode fiber components. Applied Optics, 21(15): 2671, 1982. DOI: 10.1364/AO.21.002671
[5] C.-H. Chen, M. Chen, Y.-H. Lin. An electrically-tunable liquid crystal lens coupler for the fiber communication systems. Proceedings of SPIE, 9384 (Emerging Liquid Crystal Technologies X): 938411, 2015. DOI: 10.1117/12.2081973
[6] D.-K. Yang, S.-T. Wu. Fundamentals of Liquid Crystal Devices, 2nd ed. John Wiley & Sons, 2014. ISBN 978-1-118-75200-5. DOI: 10.1002/9781118751992
[7] A. N. Simonov, G. Vdovin, M. Loktev. Liquid-crystal intraocular adaptive lens with wireless control. Optics Express, 15(12): 7468, 2007. DOI: 10.1364/OE.15.007468
[8] J. Li, S. Gauza, S.-T. Wu. Temperature effect on liquid crystal refractive indices. Journal of Applied Physics, 96(1): 19–24, 2004. DOI: 10.1063/1.1757034
[9] M. Born, E. Wolf. Principles of Optics, 7th ed. Cambridge University Press, 1999. DOI: 10.1017/CBO9781139644181
[10] R. J. Noll. Zernike polynomials and atmospheric turbulence. Journal of the Optical Society of America, 66(3): 207, 1976. DOI: 10.1364/JOSA.66.000207
[11] A. F. Naumov, M. Yu. Loktev, I. R. Guralnik, G. Vdovin. Liquid-crystal adaptive lenses with modal control. Optics Letters, 23(13): 992, 1998. DOI: 10.1364/OL.23.000992
[12] G. D. Love. Wave-front correction and production of Zernike modes with a liquid-crystal spatial light modulator. Applied Optics, 36(7): 1517, 1997. DOI: 10.1364/AO.36.001517
[13] J. W. Goodman. Introduction to Fourier Optics, 3rd ed. Roberts & Company, 2005. ISBN 978-0-9747077-2-3
[14] R. K. Tyson. Principles of Adaptive Optics, 4th ed. CRC Press, 2015. ISBN 978-1-4822-5233-0. DOI: 10.1201/b19712
[15] W. Yang, F. Knorr, J. Popp, I. W. Schie. Development and evaluation of a hand-held fiber-optic Raman probe with an integrated autofocus unit. Optics Express, 28(21): 30760, 2020. DOI: 10.1364/OE.401207
[16] S. Murali, K. S. Lee, J. P. Rolland. Invariant resolution dynamic focus OCM based on liquid crystal lens. Optics Express, 15(24): 15854, 2007. DOI: 10.1364/OE.15.015854